6. SERVICE
6.1. Fault identification and repair
In case of any difficulties during actuation of the device or during its exploitation, please return to these chapters in this
manual, which explain the performed operation. This aims to ensure that the device is properly operated. If you still
experience difficulties, the following hints will help you solve the problem.
The device is not working... – Make sure that:
• The device is connected to the supply network
• Voltage and frequency in the network are compliant with those recommended by the producer, 230V/50Hz
• The main switch is turned on
• Thermostat is turned on (This concerns the Igloo thermostat – If only two spots are visible on the display – turn on
the thermostat)
Water leaks from underneath or inside the chamber
•
Check the leveling accuracy of the machine
•
Check the drainage of the drain hoses
•
Empty the drip tray or drip container
•
Verify that the trough and the evaporator do not have large amounts of ice - defrost, if necessary
The device is operating, but the lighting is off...– Make sure that:
• Lighting switch is turned on
• Fluorescent lamp or starting switch of the device are not burnt
The device does not reach the proper temperature, the lighting is on...– Make sure that:
• The main switch is on
• Temperature setting on the thermostat is properly set
• Thermostat works properly
• The condenser is clean, if necessary – clean the condenser
• Ambient temperature does not exceed 25ºC
• Enough time has passed for products to be cooled
• Ventilation holes of the device are not blocked
(This concerns the “IGLOO” thermostat) thermostat displays C0 or C1 or C2 instead of displaying temperature:
This situation shall occur, when one of temperature regulation sensors has been destroyed. The following
messages may be displayed in such case:
• C0 – temperature sensors inside the chamber are damaged – call authorized service
• C1 – failure of evaporator sensor - call authorized service
• C2 – failure of condenser alarm sensors (or failure of second evaporator sensor) – call authorized service
(Dotyczy term. „CAREL”) Termostat wyświetla E0 lub E1 lub L0 lub HI lub EE lub Ed lub DF zamiast
temperatury:
• E0 – failure of temperature sensor inside the chamber – call authorized service
• E1 – failure of evaporator sensor – call authorized service
• L0 – low temperature alarm (lower than temperature range set within the device – call authorized service
• HI – high temperature alarm – call authorized service
• EE – internal defect of the regulator – call authorized service
• Ed – max. defrosting time exceeded
• DF – defrosting in progress (this is not an alarm signal)
Cleaning the condenser requires removing the windchest 1. Clean the condenser fins 2 with a soft brush. It is advisable
to use a vacuum cleaner or compressed nitrogen to suck/blow out the dirt between the fins with in case of heavy dirt (fins
clamping).
Содержание INNOVA T 74
Страница 39: ...ru 39 INNOVA www igloo pl 1 1 2 3 70 4 5 6 7 8 9 10 2 1 2 3 4 v 7 47 48...
Страница 41: ...ru 41 INNOVA www igloo pl 4 10 3 3 1 PBUE 3 2 3 41 1 40 C 2 8 2 42...
Страница 42: ...ru 42 INNOVA www igloo pl 2 4 1 Igloo Carel 7 47 48 D 4 2 10 C 15 C 40 90 4...
Страница 43: ...ru 43 INNOVA www igloo pl 5 5 5 1...
Страница 44: ...ru 44 INNOVA www igloo pl 11 46 6 6 1 230 50 Igloo 25 C...
Страница 46: ...ru 46 INNOVA www igloo pl 6 2 6 1 Igloo 15 50 NS 1 18 39 INNOVA 7 7 1 IGLOO 7 Igloo 2 1 3 4 6 5...
Страница 47: ...ru 47 INNOVA www igloo pl 1 2 3 4 5 6 3 3 2 www igloo pl 7 2 CAREL 8 Carel 1 2 3 4 5...
Страница 48: ...ru 48 INNOVA www igloo pl 1 2 4 4 5 1 5 1 www alfaco pl IGLOO...