4.2 Example configuration 2
Unit
LK8123 (probe length L= 472 mm)
Medium to be detected
Coolant emulsion
Operating mode
Automatic medium detection (
→ 5.2.1
)
Installation environment Metal tank, installation to Fig. 4-2
►
Install unit.
►
Observe the distances (x), (u) and (c):
x
min. 4.0 cm
u
min. 1.0 cm
c
max. 23.0 cm
►
Ground sensor and tank via an electrical
connection
(→ 7)
.
►
Observe the maximum permitted level (b).
A distance (a2) greater than 5.0 cm must be
observed between maximum level (b) and
mounting element.
►
Observe the parameter setting sequence:
-[MEdI] = [Auto]
(→ 10.2.3)
-[OFS] = (u); e. g. (u) = 1.0 cm
(→ 5.2.4)
-[SP1] = Set the switch point at a distance (a2)
Fig. 4-2
c
b
u
a2
x
Adjustable step increment: 0.5 cm.
Switch point [SP1] is used as overflow prevention (pump off, close inlet,
...).
►
Unit must be reinitialised:
►
Switch the operating voltage off and on again.
> The unit is ready for operation.
►
Make further settings if necessary.
►
Check whether the unit operates correctly.