background image

9

RS-925D

9

Maintenance

5.3.2  General Considerations

Before starting the calibration procedure, you need 

to consider the following:

• 

Calibration environment should be 23°C and 

less than 50% relative humidity.

•  Test instruments should be sufficiently more 

accurate than the RS-925D unit, and/or the 

uncertainty of the measurement instrumenta-

tion has to be considered in the calibration 

Test Uncertainty Ratio (TUR).

• 

The testing equipment and the  RS-925D unit 

should stabilize at laboratory conditions for 

at least 24 hours.

• 

Kelvin  type  4-wire  test  leads  should  be 

used  to  obtain  accurate  low  resistance 

measurements.

• 

Steps should be taken to minimize thermal 

emf effects, such as using a meter with “True 

Ohm” capacity.

• 

Accepted  metrology  practices  should  be 

followed.

5.3.3  Required Equipment

Many combinations of standards, transfer standards, 

meters, and bridges may be used to calibrate this in-

strument.  The following are some possible choices:

• 

Resistance Standards or Transfer Standards 

for 1 

, 10 

, 100 

, 1 k

, 10 k

, 100 k

1 M

, and 10 M

 per step, calibrated to 

±10 ppm.  IET options include the follow

-

ing models:

• 

HATS-LR

• 

HATS-Y

• 

SRL Series

The 1 

, and 10 

 transfer standards are 

optional, and are only used to take advantage, 

if desired, of the trimmability of these two 

decades

• 

Precision  resistance  measurement  bridge 

or  multimeter,  with  a  transfer  accuracy 

of ±l ppm.  Options include:

• 

Guildline Model 9975

• 

Measurements International Model 

6000A

• 

ESI  model  242,  242A,  242C,  or 

242D

• 

A high-precision, high-stability digi-

tal multimeter (e.g. Fluke 8508A)

along with a set of resistance stan-

dards for ratio mode.

5.3.4  Calibration procedure

To calibrate the RS-925D unit, proceed as follows:

1.  Set up the calibration equipment in the re-

sistance measurement mode and exercise the 

switches 10 times in each direction.

2.  Allow the switches to cool for 15 minutes.

3. 

Confirm the minimum resistance of the unit.

Allow a confidence band for the uncer

-

tainty of the measuring instrument and 

setup.

4.  Determine the allowable upper and lower 

limits  for  each  resistance  setting  of  each 

decade based on the specified accuracy and 

the confidence band.

For the 

 RS-925D 

series, the limits for 

any resistance “R” are:

 

[R±(20 x 10

-6

R + 0.0005Ω)]. 

5. 

Confirm that the resistances fall within these  

limits.

6.  If any resistances fall outside these limits, 

the associated switch assembly may require 

service or replacement.

Содержание RS-925D

Страница 1: ...www ietlabs com IET LABS INC Email info ietlabs com Precision instruments for test and measurement RS 925D Resistance Standard User and Service Manual Copyright 2014 IET Labs Inc Visit www ietlabs com...

Страница 2: ...ruments for test and measurement IET Model RS 925D This RS 925D is the IET Labs version of the original esi RS 925D Decade Resistance Standard The IET unit is updated and features a number of signific...

Страница 3: ...i...

Страница 4: ...ii...

Страница 5: ...ermal emf Considerations 5 4 3 Dial Setting 6 4 4 Rheostat Operation 6 4 5 Power Considerations 6 4 6 Environmental Conditions 7 4 7 Switch Conditioning 7 4 8 Meter Shunt Applications 7 4 9 Kelvin Bri...

Страница 6: ...al Operating Guide Affixed to Unit 3 Figure 4 1 Rheostat Dial 6 Figure 4 2 Rheostat Schematic 6 Figure 4 3 Kelvin Bridge Connections 7 Table 5 1 Trimming Potentiometers 11 Figure 5 1 Typical Trimmer B...

Страница 7: ...e reading The RS 925D Resistance Standard employs com pletely enclosed dust tight very low contact resistance switches They feature solid silver alloy contacts and quadruple leaf silver alloy wipers w...

Страница 8: ...1 10 ppm 5 ppm typical 10 k 100 k 10 mA 5 W 3 0 1 100 k 1 M 3 mA 2 000 V peak 3 0 1 Wiring and switch resistance NA 50 C 0 2 W NA At 23 C true ohm measurement 30 70 RH absolute reading SI traceable N...

Страница 9: ...r Decade positions Resistor type whichever applies fi rst 100 division 20 resolution 10 m 2 A NA 20 1 20 ppm 0 5 m Continuous Rheostat 1 m 10 m 2 A NA 20 1 11 positions 0 10 Resistance wire 10 m 100 m...

Страница 10: ...ard temperatures near 23 C 3 3 Repackaging for Shipment If the instrument is to be returned to IET Labs contact the Service Department at the number or address shown on the front cover of this manual...

Страница 11: ...sistance or low resistance increments are important care must be taken in connecting to the terminals of the decade box In particular to keep contact resistance to a mini mum 4 terminal connections sh...

Страница 12: ...etting in the next decade into consideration To get 1 000 for example the switches should be set to show 0 9 10 0 Since the highest decade has an additional 11 po sition resistance values of over 1 2...

Страница 13: ...connected to the current source 2 Determine the current by the ratio of the mea sured voltage to the known resistance Using the 4 terminal technique helps avoid errors caused by the voltage drops in t...

Страница 14: ...e housing 2 Remove any dust or debris using optical grade dry compressed air or a clean brush 3 Replace the housing and reattach the 4 hous ing screws The front panel should be periodically cleaned to...

Страница 15: ...follow ing models HATS LR HATS Y SRL Series The 1 and 10 transfer standards are optional and are only used to take advantage if desired of the trimmability of these two decades Precision resistance me...

Страница 16: ...TION The RS 925D front panel should be connected to the test instrument s guard point 5 4 2 Trimming Procedure To trim any resistances proceed as follows 1 Stabilize RS 925D at laboratory temperature...

Страница 17: ...er Designation All decades but 1 k Potentiometer Designation 1 k decade 1 T0 T1 2 T1 T2 3 T2 T3 4 T3 T4 5 T4 T5 6 T5 T6 7 T6 T7 8 T7 T8 9 T8 T9 10 T9 T10 11 T10 1 M steps only Table 5 1 Trimming Poten...

Страница 18: ...HARS 4000 LX 0 1 100 m step Decade Switch Assembly Not Shown HARS 4000 LX 1 1 step Decade Switch Assembly Not Shown HARS 4000 LX 10 10 step Decade Switch Assembly Not Shown HARS 4000 LX 100 100 step...

Страница 19: ......

Отзывы: