background image

8

RS-925D

8

Maintenance

Chapter 5 

MaInTEnanCE

5.1  Maintainability and Reliability

It is possible to maintain Model RS-925D indefinitely.  

It  is  reliable  due  to  its  closed  design  and  sealed 

switches and resistors.  It is possible to adjust the unit, 

if necessary, because it has 6 trimmable decades.  The 

unit is resistant to electromagnetic interference (EMI) 

because of its metal enclosure.

5.2  preventive Maintenance

Keep the unit in a clean environment.  This will help 

prevent possible contamination.

The Model  RS-925D is packaged in a closed case and 

uses completely sealed switches. This limits the entry 

of contaminants and dust to the inside of the switches.  

If it is maintained in a clean or air-conditioned envi-

ronment, cleaning will seldom be required.  

Should cleaning be needed:

1.  Remove the 4 housing screws from the side 

of the instrument, and remove the housing.

2.  Remove  any  dust  or  debris  using  optical 

grade dry compressed air or a clean brush.

3.  Replace the housing and reattach the 4 hous-

ing screws.

The front panel should be periodically cleaned to 

eliminate any leakage paths from near or around the 

binding posts.  To clean the front panel:

Wipe the front panel clean using alcohol and a 

lint-free cloth.

5.3  Calibration

The Model RS-925D may be employed as a stand-

alone instrument or as an integral component of a 

system.  If used as part of a system, it should be 

calibrated as part of the overall system to provide an 

optimum system calibration.

If the RS-925D is employed as a stand alone device, 

the following should be observed:

• 

Calibration Interval

• 

General Considerations

• 

Required Equipment

• 

Calibration Procedure

5.3.1  Calibration Interval

The recommended RS-925D Series calibration inter-

val is twelve (12) months.  

If the instrument is used to transfer resistance values 

only, recalibration is not required, assuming that there 

has been no drastic change in the deviations of any 

individual resistors.

Содержание RS-925D

Страница 1: ...www ietlabs com IET LABS INC Email info ietlabs com Precision instruments for test and measurement RS 925D Resistance Standard User and Service Manual Copyright 2014 IET Labs Inc Visit www ietlabs com...

Страница 2: ...ruments for test and measurement IET Model RS 925D This RS 925D is the IET Labs version of the original esi RS 925D Decade Resistance Standard The IET unit is updated and features a number of signific...

Страница 3: ...i...

Страница 4: ...ii...

Страница 5: ...ermal emf Considerations 5 4 3 Dial Setting 6 4 4 Rheostat Operation 6 4 5 Power Considerations 6 4 6 Environmental Conditions 7 4 7 Switch Conditioning 7 4 8 Meter Shunt Applications 7 4 9 Kelvin Bri...

Страница 6: ...al Operating Guide Affixed to Unit 3 Figure 4 1 Rheostat Dial 6 Figure 4 2 Rheostat Schematic 6 Figure 4 3 Kelvin Bridge Connections 7 Table 5 1 Trimming Potentiometers 11 Figure 5 1 Typical Trimmer B...

Страница 7: ...e reading The RS 925D Resistance Standard employs com pletely enclosed dust tight very low contact resistance switches They feature solid silver alloy contacts and quadruple leaf silver alloy wipers w...

Страница 8: ...1 10 ppm 5 ppm typical 10 k 100 k 10 mA 5 W 3 0 1 100 k 1 M 3 mA 2 000 V peak 3 0 1 Wiring and switch resistance NA 50 C 0 2 W NA At 23 C true ohm measurement 30 70 RH absolute reading SI traceable N...

Страница 9: ...r Decade positions Resistor type whichever applies fi rst 100 division 20 resolution 10 m 2 A NA 20 1 20 ppm 0 5 m Continuous Rheostat 1 m 10 m 2 A NA 20 1 11 positions 0 10 Resistance wire 10 m 100 m...

Страница 10: ...ard temperatures near 23 C 3 3 Repackaging for Shipment If the instrument is to be returned to IET Labs contact the Service Department at the number or address shown on the front cover of this manual...

Страница 11: ...sistance or low resistance increments are important care must be taken in connecting to the terminals of the decade box In particular to keep contact resistance to a mini mum 4 terminal connections sh...

Страница 12: ...etting in the next decade into consideration To get 1 000 for example the switches should be set to show 0 9 10 0 Since the highest decade has an additional 11 po sition resistance values of over 1 2...

Страница 13: ...connected to the current source 2 Determine the current by the ratio of the mea sured voltage to the known resistance Using the 4 terminal technique helps avoid errors caused by the voltage drops in t...

Страница 14: ...e housing 2 Remove any dust or debris using optical grade dry compressed air or a clean brush 3 Replace the housing and reattach the 4 hous ing screws The front panel should be periodically cleaned to...

Страница 15: ...follow ing models HATS LR HATS Y SRL Series The 1 and 10 transfer standards are optional and are only used to take advantage if desired of the trimmability of these two decades Precision resistance me...

Страница 16: ...TION The RS 925D front panel should be connected to the test instrument s guard point 5 4 2 Trimming Procedure To trim any resistances proceed as follows 1 Stabilize RS 925D at laboratory temperature...

Страница 17: ...er Designation All decades but 1 k Potentiometer Designation 1 k decade 1 T0 T1 2 T1 T2 3 T2 T3 4 T3 T4 5 T4 T5 6 T5 T6 7 T6 T7 8 T7 T8 9 T8 T9 10 T9 T10 11 T10 1 M steps only Table 5 1 Trimming Poten...

Страница 18: ...HARS 4000 LX 0 1 100 m step Decade Switch Assembly Not Shown HARS 4000 LX 1 1 step Decade Switch Assembly Not Shown HARS 4000 LX 10 10 step Decade Switch Assembly Not Shown HARS 4000 LX 100 100 step...

Страница 19: ......

Отзывы: