Troubleshooting
107
EN
Error message
Possible cause
Measures
The evaluation unit is not
reacting to data transfer but
co
is shown on the display.
1. Data transmission cable has not
been inserted properly into the
PC.
1. Check whether the 9-pin plug of
the data transmission cable is
securely located in the device’s
interface socket.
2. Also see Err 4
2. Also see Err 4
The Mobil-O-Graph
®
takes
measurements every two
minutes.
Log 9 is set in the Mobil-O-
Graph
®
.
Set log 1 or 2 (see chapter 3.4.5
“Specifying the desired
measurement log”)
The desired log cannot be
set with the button
combination.
There are still readings from the
last patient in the memory
Delete the data in the blood
pressure monitor but make sure
that the data has been saved (see
chapter 3.4.2 “Clearing the
memory”)
The blood pressure monitor
cannot be switched on.
1. The power packs or batteries
have been incorrectly inserted
1. Reinsert both power packs or
batteries and ensure correct
polarity.
2. The power pack or battery
voltage is too low
2. Replace the power packs or
batteries
3. The display is faulty
3. Send the device to your
specialist supplier or directly to IEM
GmbH for repair.
Содержание Mobil-O-Graph NG
Страница 2: ...Revision B 17 10 2018 ...
Страница 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG ...
Страница 66: ...64 ...
Страница 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG ...
Страница 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG ...
Страница 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG ...
Страница 254: ...252 ...
Страница 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG ...
Страница 320: ...318 ...
Страница 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG ...
Страница 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden ...
Страница 364: ...Probleemoplossing 362 Foutmelding Mogelijke oorzaak Maatregelen vergelijk deze met de manchetopdruk ...
Страница 384: ......
Страница 385: ......