Troubleshooting
106
Error message
Possible cause
Measures
Err 7
The memory of the blood pressure
measuring device is full (a
maximum of 300 measurements
and events can be stored or in the
case of PWA a maximum of 260)
Delete the data in the blood
pressure monitor but make sure
that the data has been saved (see
chapter 3.4.2 “Clearing the
memory”).
Err 8
Measurement cancelled by
pressing a button
Repeat the measurement
Err 9 +
possible
continuous alarm
until a button is
pressed
1. Residual pressure inside the
cuff
1. Wait for the cuff to deflate
completely.
2. Zero point comparison could
not be performed successfully.
2. Send the device immediately to
your specialist supplier or directly
to IEM GmbH for checking.
Err 10 +
Continuous
alarm until a
button is pressed
1. Serious error due to pressure
build-up outside the measuring
process (pump has been switched
on against orders)
Send the device immediately to
your specialist supplier or directly
to IEM GmbH for checking and
repairs.
2. These error messages all show
a severe error in the program
code.
Содержание Mobil-O-Graph NG
Страница 2: ...Revision B 17 10 2018 ...
Страница 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG ...
Страница 66: ...64 ...
Страница 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG ...
Страница 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG ...
Страница 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG ...
Страница 254: ...252 ...
Страница 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG ...
Страница 320: ...318 ...
Страница 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG ...
Страница 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden ...
Страница 364: ...Probleemoplossing 362 Foutmelding Mogelijke oorzaak Maatregelen vergelijk deze met de manchetopdruk ...
Страница 384: ......
Страница 385: ......