Troubleshooting
105
EN
Error message
Possible cause
Measures
Err 5
bAtt
1. Power pack or battery voltage
too low
1. Replace the power packs or
batteries
2. Power packs or batteries
defective
2. The power pack or battery
voltage is correct but
bAtt
is shown
on the display during cuff inflation.
Replace the power packs
3. Battery contacts are corroded
3. Clean the battery contacts with a
cotton cloth and a little alcohol.
Err 6 +
possible
continuous alarm
until a button is
pressed
1. Build-up of air
1. Check the cuff for a build-up of
air or a kink in the tube. If there is a
kink in the cuff tube, lay the tube
out. Otherwise, send the device in
immediately.
2. Blood pressure cuff incorrectly
connected
2. Connect the cuff to the device
(see chapter 3.3.5 “Cuff
connection”)
3. Leak in the cuff or connecting
tube
3. If necessary, replace the cuff or
connecting tube.
Содержание Mobil-O-Graph NG
Страница 2: ...Revision B 17 10 2018 ...
Страница 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG ...
Страница 66: ...64 ...
Страница 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG ...
Страница 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG ...
Страница 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG ...
Страница 254: ...252 ...
Страница 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG ...
Страница 320: ...318 ...
Страница 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG ...
Страница 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden ...
Страница 364: ...Probleemoplossing 362 Foutmelding Mogelijke oorzaak Maatregelen vergelijk deze met de manchetopdruk ...
Страница 384: ......
Страница 385: ......