Manufacturer’s EMC guidelines
51
EN
Electromagnetic immunity
The
eConnect
is designed for operation in an electromagnetic environment as specified below. Only use the
eConnect
in such an environment.
Measurement of
interference emissions
IEC 60601-1 test level
Compliance level
Electromagnetic
environment
– guidelines
Electrostatic discharge
(ESD) in accordance with
IEC 61000-4-2
±/- 8kV contact discharge
±/- 15kV air discharge
±/- 8kV contact discharge
±/- 15kV air discharge
Floors should consist of
wood or cement or ceramic
tiles. If the floor consists of
synthetic materials, relative
humidity must be at least
30%.
IEC 61000-4-4
Not applicable
IEC 61000-4-5
Not applicable
IEC 61000-4-11
Not applicable
Magnetic field at supply
frequency (50/60 Hz) in
accordance with IEC 61000-
4-8
30 A/m
30 A/m
Magnetic fields at mains
frequency should match the
typical values found in
business and hospital
environments.
Radiated RF interference
in accordance with IEC
61000-4-3
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1
kHz Sinus)
PM; 18 Hz
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1
kHz Sinus)
PM; 18 Hz