Operator Manual ilex with IPS500A
L280-0688 REV C
Recommended separation distances between portable and mobile RF communications
equipment and the ILEX system
The ilex system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ilex system can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the ilex system as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
__
d = 1.2
√
P
80 MHz to 800 MHz
__
d = 1.2
√
P
800 MHz to 2,5 GHz
__
d = 2.3
√
P
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.20
1.20
2.30
10
3.79
3.79
7.27
100
12.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.