Operator Manual ilex with IPS500A
L280-0688 REV C
12.
Specifications
Mode of Operation
▪
For continuous use
Type of Equipment
:
▪
Class I (as defined by IEC 60601-1, 3
rd
Edition
▪
CENELEC EN 60601-1
▪
ANSI/AAMI ES60601-1
▪
IEC60601-1 3
rd
Edition
Electrical:
▪
Supply Voltage:
▪
120±5% Vac 60Hz through NEMA 5-15P Hospital Grade Plug
▪
230±10% Vac 50Hz through country specific plug
▪
Output Voltage: 120 VAC 60Hz (1800VA) or 230 VAC 50 Hz (1800VA).
▪
Current Rating: Less than 12 Amps @ 120 Vac or 7 Amps @ 230 Vac
▪
Isolation transformer limits leakage current to 50µ
▪
Isolation transformer is certified to meet Medical Safety Standards
ANSI/AAMI ES60601-1
Environmental
:
▪
Operating Temperature Range: 32
F to 104
F (0
C to 40
C)
▪
Operating Humidity Range: 10% to 80% relative humidity, noncondensing.
▪
Operating Pressure Range: 700 hPa to 1060 hPa.
▪
Transport & Storage Temperature Range: -4
F to 140
F (-20
C to +60
C).
▪
Transport & Storage Humidity Range: 10% to 80% relative humidity, noncondensing.
▪
Transport & Storage Pressure Range: 700 hPa to 1060 hPa.
▪
NOT suitable for oxygen rich environment
▪
NOT suitable for flammable anesthetic mixtures