Operator Manual ilex with IPS500A
L280-0688 REV C
19in x 28in
[48cm x 71cm]
20in x 20in
[50cm x 50cm]
12in
31cm
9in
22cm
16in
40cm
92in
72in
233cm
182cm
3012"
77cm
31"
790mm
55in
140cm
ACROSS
HANDLES
3534"
910mm
ilex55 Mobile Video Display System
IMAGE DIAGNOSTICS, INC.
Power Switches
In the event power switches have been toggled “OFF”, please use the following to turn power on
to the different components. To power up the monitor there is a switch located on the rear of the
monitor where video inputs are. To power up the IPS500A reach behind the tray on the top shelf
on the right side of the box and there is a power on/off switch. Under normal use these buttons
can stay in the “ON” position.
On/off Button, use the green toggle switch/thermal breaker to turn the system on
and off and to reset the breaker.
Up/Down switch, use the monitor elevation toggle switch
to adjust monitor height.
When the red emergency stop button is pushed in, elevation of the monitor is not allowed.
Always push the red emergency stop button when moving the ilex system to
reduce the risk of damage to the unit
. Reset by rotating the button a quarter turn.
Maximum duty cycle of the actuator is 10% or 2 minutes of continuous
use followed by 18 min. not in use. Exceeding this limit could damage
the device and require replacement.
Monitor (Rear of unit)
Master
On/Off
IPS500A Video Router
(Rear of unit)