IDEXX VetLab
UA
Analyzer Operator’s Guide
Components
How it Works
A prepared IDEXX UA Strip is placed on the test strip tray, which moves it into the analyzer and
under the reading head. The analyzer reads the gray reference pad first, and then reads each of
the test pads on the strip. Results are available in approximately 80 seconds.
IMPORTANT:
Only IDEXX UA Strips can be used with the IDEXX VetLab UA Analyzer.
The analyzer contains light-emitting diodes (LEDs) that emit light at various wavelengths. Reading
is done electro-optically, as follows:
LED
Test pad
Test strip
Phototransistor
A/D converter
(analog to digital)
Microprocessor
Semi-quantitative result
Cover:
Press here to open the cover, load
printer paper, or load a software
upgrade code plug
Display menu
Function keys
Test strip tray
Start button:
Starts the reading process
NOTE:
Press the Start button
during any submenu process to
return the analyzer to ready-to-
measure mode.
On/Off switch
Serial interface:
for connection
to the IDEXX VetLab
Station
Power socket:
connects the analyzer
to the AC power
adapter
5-pin DIN socket
(not used)
Printout:
optional if connected to the
IDEXX VetLab* Station
IDEXX UA Strip