background image

11

To Change the Date Settings

1.  At the bottom of the Home screen, tap 

More

, tap 

Settings

, and then tap 

Date

.

2.  If you want to change the date format, tap the current format (underneath the current date 

display) and then select a new date format. 

3.  Use the numeric keypad to change the date, if necessary.
4.  Tap the 

Home

   icon to return to the Home screen.

To Change the Time Settings

1.  At the bottom of the Home screen, tap 

More

, tap 

Settings

, and then tap 

Time

.

2.  If you want to change the time format, select a new format (underneath the current time 

display). 

3.  Use the numeric keypad to change the time, if necessary.
4.  Tap the 

Home

   icon to return to the Home screen.

To Change the Language Settings

1.  At the bottom of the Home screen, tap 

More

, tap 

Settings

, and then tap 

Language

.

2.  Select a new language. 
3.  Tap the 

Home

   icon to return to the Home screen.

To Change the Location Settings

1.  At the bottom of the Home screen, tap 

More

, tap 

Settings

, and then tap 

Location

.

2.  Select a new location. 
3.  Tap the 

Home

   icon to return to the Home screen.

Modifying the Preferences

Содержание SNAP Pro

Страница 1: ...SNAP Pro Analyzer Operator s Guide...

Страница 2: ...industrial property rights covering this document or subject matter in this document The furnishing of this document does not give a license to these property rights except as expressly provided in a...

Страница 3: ...onnected to an IDEXX VetLab Station 8 Confirming Results 9 Viewing Test Results 9 Modifying the Preferences 10 Changing the ProRead Settings 10 Changing the Connection Settings 10 Viewing Past Notific...

Страница 4: ...he IDEXX VetLab Station helps you attain a comprehensive picture of your patient s health with the ability to view test results from multiple analyzers on a single report determine disease progression...

Страница 5: ...n snap into place 3 Connect the power cord to the analyzer and an appropriate electrical outlet The status light should appear steady yellow Note It will take up to 8 hours to fully charge the analyze...

Страница 6: ...analyzer is being used cordlessly or plugged into an electrical outlet corded When the status light appears And is The analyzer Green Solid Has interpreted that all results on the test are negative O...

Страница 7: ...power button until you hear a beep To Power the Analyzer Off Hold the power button for 3 seconds and release when you hear a beep Charging the Analyzer The analyzer will charge automatically when it...

Страница 8: ...analyzer activates the test and then a countdown timer displays When analysis is complete results are shown on the results screen and are sent to the IDEXX VetLab Station automatically Tip The color...

Страница 9: ...lts tap the Check Mark icon The results will then be saved on the analyzer and sent to the IDEXX VetLab Station if applicable Note To cancel the test and return to the Home screen tap the Close icon V...

Страница 10: ...etwork and select the IDEXX network OR Tap No if you do not want to connect the analyzer to an IDEXX VetLab Station 3 Tap the Home icon to return to the Home screen Viewing Past Notifications 1 At the...

Страница 11: ...ettings and then tap Time 2 If you want to change the time format select a new format underneath the current time display 3 Use the numeric keypad to change the time if necessary 4 Tap the Home icon t...

Страница 12: ...il you the updated software for the analyzer For detailed upgrading instructions see the insert included with the upgrade USB drive Cleaning the Internal Components To Be Performed Weekly or As Needed...

Страница 13: ...ach solution reflectors lens activator axle 5 Clean the 3 white reflectors the axle and the activator using a new lint free wipe or cotton swab dampened with 70 isopropyl alcohol mild liquid soap and...

Страница 14: ...IDEXX Laboratories to obtain replacement batteries This symbol is placed in accordance with the European Union Directive 2002 96 on the Waste Electrical and Electronic Equipment the WEEE Directive If...

Страница 15: ...clude many behind the scenes improvements We also recommend rebooting the analyzer weekly If your software is up to date and rebooting did not resolve your issue consult this section for more informat...

Страница 16: ...e always upgrade the software when prompted Improper sample preparation An anticoagulant was not used with the whole blood sample The pipette provided with the SNAP test was not used to properly measu...

Страница 17: ...use The analyzer base isn t properly locked into place ensure the base is secure and that both fasteners on the base are in the locked position The batteries are in backwards or missing see photo D sh...

Страница 18: ...sumption Operating maximum 5 W Batteries Main Batteries Output 3 7 V DC 2 600 mAh Chemistry Lithium ion Type Portable 18650 Rechargeable Yes Weight 50 0 g Button Cell Battery Output 3 V DC 48 mAh Chem...

Страница 19: ...n the European Union and is strongly encouraged in all other countries If the equipment is used in a manner other than specified the protection provided by the analyzer may be impaired The SNAP Pro An...

Страница 20: ......

Отзывы: