#61-096

200A~

61-096

CAT. IV

600V~

VA

H

Lo ZI

Ω

AUTO SEL

AC

DC

HOLD

VOLTS: 0-750VAC

0-1000VDC

AMPS: 0-200AAC

OHMS:  0-2000

Continuity

Split-Jaw

Smart-Meter

Ω

COM

CAT. IV

600V

CAT. III

1000V

V-

Ω

Split-Jaw™

Smart-Meter

Register your product and 

access more information at 

www.idealindustries.com

 

Read First: Safety 

Information

 

Understand and follow operating 

instructions carefully. Use the meter, 

test leads and all accessories only 

as specified in this manual; 

otherwise, the protection provided 

by the meter can be impaired.

                       

WARNING

To avoid possible electric 

shock, personal injury or death, 

follow these guidelines:

•  Do not use if meter appears damaged. Visually inspect the 

meter to ensure case and jaws are not cracked.

•  Inspect and replace test leads if insulation is damaged, 

metal is exposed, or probes are cracked. Pay particular 

attention to the insulation surrounding the connectors.

•  Always ensure the meter, test leads and all accessories 

meet or exceed the measurement category required in the 

working environment. (i.e. CAT rating)

•  Note that the measurement category and voltage rating 

of combinations of the meter, the test leads, and the 

accessories is the lowest of the individual components. 

•  Do not use meter if it operates abnormally as protection 

maybe impaired.  

•  Do not use during electrical storms or in wet weather.

•  Do not use around explosive gas, dust, vapor, amperage 

or in damp or wet environments.

•  Do not apply more than the rated voltage to the meter. 

•  Remove the test leads from the input jacks before 

measuring current.

•  Replace battery as soon as battery indicator         appears 

to avoid false readings.

•  Remove the test leads from the meter prior to removing 

battery cover.

•  Do not use without the battery and battery cover properly 

installed.

•  Do not attempt to repair this unit as it has no user-

serviceable parts.

•  Use the proper terminals, functions and range for your 

measurements.

•  Never ground yourself when taking electrical measurements.

•  Connect the black common lead to ground or neutral 

before applying the red test lead to potential voltage.  

Disconnect the red test lead from the voltage first.

•  When using the probes, keep fingers behind the guard 

rings of the probe tips.

•  Voltages exceeding 30VAC or 60VDC pose a shock 

hazard so use caution.

          

CAUTION

To protect yourself, think “Safety First”:

•  Comply with local and national safety codes. 

•  Use appropriate personal protective equipment such as 

face shields, insulating gloves, insulating boots, and/or 

insulating mats.

•  Before each use:

  -  Perform a continuity test by touching the test leads 

    together to verify the functionality of the battery and 

    test leads.

  -  Use the 3 Point Safety Method. (1) Verify meter 

    operation by measuring a known voltage. (2) Apply 

    meter to circuit under test.  (3) Return to the known 

    live voltage again to ensure proper operation.

•  Always work with a partner.

~

...

Symbols on the Unit

Risk of Danger. Important Information. See 

Manual.

Hazardous voltage.

Application around and removal from 

Hazardous Live conductors is permitted.
AC (Alternating Current)
DC (Direct Current)

Earth Ground

CAT II

Measurement Category II applies to test 

and measuring circuits connected directly 

to utilization points (socket outlets and 

similar points) of the low-voltage MAINS 

installation

CAT III

Measurement Category III applies to mea-

suring circuits connected to the distribution 

part of the building’s low-voltage MAINS 

installation

CAT IV

Measurement Category IV applies to test 

and measuring circuits connected at the 

source of the building’s low-voltage MAINS 

installation

Содержание SPLIT-JAW 61-096

Страница 1: ...attery cover Do not use without the battery and battery cover properly installed Do not attempt to repair this unit as it has no user serviceable parts Use the proper terminals functions and range for...

Страница 2: ...Off On HOLD HOLD HOLD AUTO SEL Press AUTO SEL HOLD HOLD HOLD Press 2sec COM V AUTOMATIC TESTER HOLD V AC 200A ACV 1 0V CAT IV 600V AUTO SEL AUTO SEL V AC Voltage AC DC Resistance Ohms DCV Measuring Ra...

Страница 3: ...formed by qualified personnel Cleaning Periodically wipe the case with a dry cloth and detergent Do not use abrasives or solvents Priority of Function 100A 200A CAT IV 600V 200A CAT IV 600V HOLD HOLD...

Страница 4: ...nput voltage to approximately 200K at 600V Resistance Continuity Function Range Accuracy 0 0 99 9 2 1 100 2000 2 2 dgt Temperature coefficient multiplies by 1 5 when operating temperature is 40 C 50 C...

Страница 5: ...a de la bater a No use el mult metro sin la bater a ni sin la tapa de la bater a correctamente instalada No intente reparar esta unidad ya que no tiene piezas reparables por el usuario Use las termina...

Страница 6: ...LD HOLD HOLD Press 2sec COM V AUTOMATIC TESTER HOLD V AC 200A ACV 1 0V CAT IV 600V AUTO SEL AUTO SEL V AC Apagado Apagado Encendido Encendido Pulse 30 minutos Pulse tiempo 2 segundos Voltaje de CA CC...

Страница 7: ...ioridad de las funciones 100A 200A CAT IV 600V 200A CAT IV 600V HOLD HOLD HOLD AUTOMATIC TESTER AUTOMATIC TESTER A AUTO SEL Press AC A AUTO SEL AC H A AUTO SEL AC H H A AUTO SEL AC La pantalla de cris...

Страница 8: ...e contaminaci n 2 Compatibilidad electromagn tica EMC Cumple con la norma EN 61326 1 Impacto y vibraci n Vibraci n sinusoidal de acuerdo con la norma MIL T 28800E 5 a 55 Hz 3g m ximo Protecci n contra...

Страница 9: ...r sent manuel d faut la protection offerte par le multim tre pourra tre compromise AVERTISSEMENT Pour viter tout risque d lectrocution de l sions personnelles ou de mort se conformer aux directives su...

Страница 10: ...e tension connue 2 Appliquer le multim tre au circuit en cours de contr le 3 Retourner la tension active connue pour v rifier le bon fonctionnement Travailler toujours avec un quipier Symboles figuran...

Страница 11: ...CAT IV 600V AUTO SEL AUTO SEL V AC 200A AUTO SEL HOLD CAT IV 600V COM V AUTOMATIC TESTER Plage de mesure 0 2000 Continuit Courant C A Amp res La m choire accepte les c bles cu jusqu au calibre 1 0 0 5...

Страница 12: ...jusqu 6000 Fr quence de mesure 5 fois s Affichage de d passement OL est affich pour la fonction la valeur r elle est affich e pour les fonctions A et V Coupure automatique de l alimentation Environ 30...

Страница 13: ...unit d fectueuse suite la v rification du d faut ou du dysfonctionnement Cette garantie ne s applique pas aux fusibles aux piles ou aux dommages r sultant d une utilisation abusive de la n gligence d...

Отзывы: