UniPRO
152810
User Guide
Page 39
The values permitted for the Limits shown on Fig 48B are:
Max Latency – 0 to 10,000,000 µs.
Max Jitter – 0 to 2,000,000 µs.
Frame Loss Ratio 0 to 1.
Frame Loss Count – 0 to 1,000,000,000.
Results
Fig 49 shows SLA Tick test result screens. The screen, ‘A’, confirms the target and that the test is
Passed, i.e. is below all limits and meets all criteria specified at set up. Select the buttons shown on
screen ‘B’, and test results are shown on the display as a series of graphs and tables. MORE (F2) toggles
between the two screens.
A
B
Fig 49
SLA Frame Error, Rx Tx Frame Counts, Rx Tx Layer Payload, Rx Frame Size, Rx Frame Types, Rx
Errors and Rx Errored Time are shown as tables only, all other results are shown as tables and graphs.
For each of the graphs (Latency results are shown at Fig 50), the x-axis represents each of the eight
services, the y-axis records the quality being measured. TABLES (F3) present the same information but
values are shown in numerals.
A
B
Fig 50