Chapter 3
Structured Cable Field Calibration and Testing
3-6
Step 3
¾
Remove the RJ45 mini patchcord from the RH and the hybrid patchcord.
¾
With the DH hybrid patchcord open-ended, select START from the DH Field
Calibration screen to perform the third calibration process.
Step 4
¾
Re-connect the RJ45 plug of the first hybrid patchcords to the Remote Handset
(RH).
¾
With the RH hybrid patchcord open-ended, press AUTOTEST from the RH to
perform the fourth calibration process. Upon completion of the calibration
process, the LANTEK handsets and the hybrid patchcords are ready for use in
testing cables.
FIELD CALIBRATION - BLOCK TESTING 66/110/BIX
This Field Calibration is a 4-step process. Steps 1 and 2 are performed with a
patchcord connected to the Handsets. Steps 3 and 4 are performed with open-ended
patchcords (only one end connected to each Handset).
For Block 66 Systems
The RJ45 block calibration adapters connected to the block during calibration is either
the T568A or 568B RJ45/66 Block adapter. The T568A adapter is used with patchcords
connected in a TIA-568A configuration. The T568B adapter is used with patchcords
connected in a TIA-568B configuration. Both block adapters have the standard RJ45
plug at the opposite end.