27
To choose the measurement parameters:
1) Use the up
[
5
]
and down
[
6
]
arrow buttons to highlight
PARAM
and press the enter
[ENTER]
button. This brings up a grid showing all of the available instantaneous parameters.
2) The selected parameters will be highlighted in black.
3) Use the up
[
5
]
, down
[
6
]
, left
[
3
]
, and right
[
4
]
arrow buttons to move the cursor through the
parameters.
4) Pressing the set
[SET]
button selects, or de-selects a parameter. Pressing set while on “CLEAR
ALL“ at the bottom of the display clears all parameters.
5) After the desired instantaneous parameters have been selected, move the cursor to the upper left-
hand corner
[INST]
and press set . This will bring up a grid showing all of the available maximum
parameters. The selected parameters will be highlighted in black and
MAX
will be displayed in the
upper left-hand corner of the screen.
6) Use the up
[
5
]
, down
[
6
]
, left
[
3
]
, and right
[
4
]
arrow buttons to move the cursor through the
available maximum parameters.
7) Pressing the set
[SET]
button selects, or de-selects a parameter. Pressing set while on “CLEAR
ALL“ at the bottom of the display clears all parameters.
8) After the desired maximum parameters have been selected, move the cursor to the upper left-hand
corner
[MAX]
and press set . This will bring up a grid showing all of the available minimum
parameters. The selected parameters will be highlighted in black and
MIN
will be displayed in the
upper left-hand corner of the screen.
9) Use the up
[
5
]
, down
[
6
]
, left
[
3
]
, and right
[
4
]
arrow buttons to move the cursor through the
available minimum parameters.
10) Pressing the set
[SET]
button selects, or de-selects a parameter. Pressing set while on “CLEAR
ALL“ at the bottom of the display clears all parameters.
11) After the desired maximum parameters have been selected, move the cursor to the upper left-hand
corner
[MIN]
and press set . This will bring up a grid showing all of the available minimum
parameters. The selected parameters will be highlighted in black and
HARM
will be displayed in the
upper left-hand corner of the screen.
12) Use the up
[
5
]
, down
[
6
]
, left
[
3
]
, and right
[
4
]
arrow buttons to move the cursor through the
parameters.
13) After the desired parameters have been selected, select the desired order of harmonics. Choosing
30 selects the individual harmonics to the 30th and 50 selects the individual harmonics to the 50th.
14) Press the enter
[ENTER]
button to save your changes and return to the record menu.
15) Press the escape
[ESC]
button to return to the record menu without saving any changes.
To return to the setup menu, press either the left arrow
[
3
]
or escape
[ESC]
button.
5.2.3 Comm
Comm is the section of the menu to set the communication protocols for downloading data. During this
process, the PowerVision™ software will auto-detect the analyzer at all possible baud rates, so it is not
necessary to change these parameters, however a faster baud rate may be desired.
Communication parameters:
•
Baud rate: choose from 1200, 2400, 4800, 9600, 19200, 38400 Factory default setting is 9600
•
Parity: choose from no, odd, even Factory default setting is no
•
Length: this is set at 8 Factory default setting is 8
•
Stop bits: chose from 1, 2 Factory default setting is 1