Time Controller Series User Manual
Section 10: Appendix: DLT Command Reference
ID QUANTIQUE SA
1227 Geneva
T +41 22 301 83 71
92
Rue Eugène-Marziano 25
Switzerland
F +41 22 301 83 79
-
timestamps_count
: total number of received timestamps
Answer example:
{
"acquisitions_count": 0,
"errors": [],
"inactivity": 1.124206,
"timestamps_count": 0
}
The status reported values allows for a basic monitoring of an ongoing acquisition.
-
acquisition_count
: should be equal to number of desired records, defined by the SCPI
command
RECord:NUMber?
and usually equals 0 during the acquisition and 1 upon its
completion.
-
errors
: should be an empty list.
-
inactivity
: should not exceed a few seconds after the end of the acquisition. This would
mean that the Time Controller is not sending any data to the DLT.
-
timestamps_count
: total number of received timestamps.