LOOP: the input block receives the data from the block linked to the OUTLoop block.
Additional information about the HIREs:ERROr? returned error code:
Device
Error code
Description
All
0
No unacknowledged error.
ID900
1
Recalibration required (DEVI:SAMP:RECA).
ID900
2
Input count rate too high (above 25 MHz).
ID900
3-4
Time between two consecutive events is sometime too short.
ID900
5
Channel disabled (device reboot required).
ID1000
1001-1003
Signal rate too high or bad input threshold.
ID1000 only
•
Some errors trigger automatic corrective actions leading to event loss on the impacted
channel.
•
If the problem persists despite proper signal rate and input threshold, try to recalibrate
the device (a power-cycle might be required on some rare occurrences).
9.2.3
Start (STARt)
STARt block
Command tree
Input
Output
Description
STARt
:ENABle
{ON|OFF}
Enable/Disable Start
:ENABle?
{ON|OFF}
:COUNter?
number
Returns the number of counts over the
integration time (CYCLE mode) or since last
:RESEt (ACCUM mode)
:INTEgrationtime
time in ms
Set counter integration time (only for CYCLE
mode)
:INTEgrationtime?
time in ms
Returns integration time
:MODE
{ CYCLe|ACCUm}
Select the counter mode
:MODE?
{ CYCLe|ACCUm} Returns the counter mode
:RESEt
Reset counter to 0
:COUPling
{AC|DC}
Set AC or DC coupling
:COUPling?
{AC|DC}
:DELAy
value in ps
Set delay on start input
:DELAy?
Value
in
ps
:EDGE
{RISIng|FALLing}
Set discriminator edge
:EDGE?
{RISIng|FALLing}
STARt block
Command tree
Input
Output
Description
:THREshold
value in V
Discriminator threshold