Co
Brite
User Manual
40
Command
Read
Write
User
access
level
Stored
if
STADEF =0
[:SOURce:]STATe/?<wsp><C>,<S>,<D>,/<P>
RW
0
✓
[:SOURce:]MONitor?<wsp><C>,<S>,<D>
R
0
-
7.9
SCPI Code error definition
Error #
Description
100
Command Syntax Command error
101
Parameter out of range
102
No data in buffer
103
device not ready
104
user level for this command not sufficient
200
Command execution error
201
user level rights to execute this command not sufficient, increase using
PASS
command
250
251
252
7.10
Alarm code definition
The following table details potential alarms raised by the unit. The alarm status can be queried
using the “alarm?” command. The response is an ASCII coded unsigned 16-bit integer number,
little endian. Convert this number into binary format to retrieve th
e alarm status. Each “1” bit
represents an active alarm.
The most common responses to “
alarm?
” are:
Use the SCPI Command “
*cls
” to clear all latched alarms.
Bit #
Alarm
Condition/Description
0
Laser Temperature too high
Base temperature of laser exceeds max. threshold.
Laser is de-activated to protect Hardware
1
Interlock active
Interlock safety switch located at back of unit is
open. This prevents switching on all laser ports
2
Controller communication Failure
Main Controller cannot communicate with other
internal components of the unit
3
Laser error
One or more lasers report an error.
4
Not used
5
Not used
6
Not used
7
Not used
8
Not used
9
Not used
10
Not used
11
Not used
12
Not used
13
Not used
14
Not used
15
Not used