11
Co
Brite
User Manual
Each laser port installed on the chassis can be en- or disabled to be considered for the H/W
trigger input
using the command “TRINACT”. All considered ports are will only start to tune if a
new setting has been set via software once the trigger in port has been initiated.
Note
: Per default, “trigger tuning” for all Lasers are disabled. Use SCPI commands to alter
configuration.
Trigger OUT
provides an output signal indicating periods at which the one or more laser ports
are tuning. The polarity of the signal and the logic of the signal can be changed via Remote
control or GUI.
Each laser port installed on the chassis can be en- or disabled to be considered for the H/W
trigger output using the command “TROUTACT”. All considered ports are combined via a logical
“OR” for the tuning active true state.
Note
: Per default, “tuning active” signals of all Lasers are considered for the joint trigger
out.