background image

 

 
 

 

nGauge AFM User Manual

 

19

 

 

 

7.4 Menu Button - Advanced Controls 

 

In the top right corner of the screen, click on the button with three horizontal lines, the "menu" 

button.  

Motor Controls 

Controls the sample platform. Motor Speed adjusts how quickly the platform 
moves. 

Approach 

makes the platform move up, 

Retract 

makes the platform go 

down, clicking either 

Approach 

or 

Retract 

takes a single step, while clicking and 

holding 

Approach

 or 

Retract

 moves the platform continuously. 

PID Controller 

Proportional is the proportional gain coefficient of the PI controller. Integral is 
the integral gain coefficient of the PI controller. Setpoint is the tip oscillation 
amplitude that the controller will try to maintain. Enable/Disable turns the 
controller on or off. 

DACs 

Digital to Analog Converters (DACs) output analog voltages. These voltages 
drive the X, Y, and Z actuators. The DAC controls allow you to set the voltage 
level that is being applied to any of the actuators, including the lateral 
actuautors (X1, X2, Y1, Y2), the Coarse Z and Fine Z, and the Leveling control 
(which is integrated into the Z actuator). There is a conductive path to the AFM 
tip and the Tip control allows you to set the tip voltage. All DACs range from 0-
3.3V 

PGAs 

Programmable Gain Amplifiers (PGAs) are able to scale the voltages output by 
the DACs. The PGAs are used to adjust the XY scan size, the range of the Coarse 
and Fine Z, the amplitude of the AC driving signal on the Z actuator, and the 
amount of leveling that is added to each scan. PGA values can range from 0-
140%. 

ADCs 

Analog to Digital Converters (ADCs) read analog voltages. Clicking 

Read All

 will 

display the current voltage level of the current sensing circuits for X, Y, and Z, 
as well as the voltage levels of the amplitude and phase measurement circuits. 
The ADCs range from 0-3.3V. 

Resistances 

Measured resistances of the MEMS actuators. Click 

Resistance

 button to 

refresh measurements. 

DDS 

DDS (direct digital synthesis) section controls the drive signal frequency. To set 
a new frequency, change both start and end frequencies to the desired 
frequency.  

Safety Configurations 

Encapsulates various safety features. Safety thresholds and features may be 
modified or disabled in this section.  

Approach Low Limit

 specifies the low signal threshold that constitutes a safe 

situation for auto-approach. If the measured signal at the beginning of auto-
approach is lower than the specified value, the approach will be aborted. It is 
recommended to leave the setting at its default value of 0.7 V.  

Check Resistances

 refers to the safety routine that is executed prior to every 

Sweep

Approach

, and 

Scan

 command. If enabled, the software will measure 

resistances and check for any anomalies.  

Limit Switch

 refers to the end-stop mechanism to prevent motor from driving 

the stage too low. In the unlikely circumstance that the switch is damaged and 
you cannot retract properly, you can uncheck the limit switch feature to 

Содержание nGauge AFM

Страница 1: ...User Manual Version 1 1 Atomic Force Microscope...

Страница 2: ...3 1 Where to set it up 6 3 2 How to set it up 7 3 3 Installing the Software 9 4 0 Basic Operation 10 4 1 Software Overview 10 4 2 Preparation and Approach 11 4 3 Imaging 12 4 4 Analyzing Data 14 5 0...

Страница 3: ...ollowing warnings and precautions should be observed to ensure the safety of the user and the instrument Use only the power supply provided The driving electronics are designed to accept 7 5Vdc and an...

Страница 4: ...ng DOWN in the storage box Use tweezers to pick them up The AFM chip can be placed on a flat surface with the tip facing up The AFM chip can be picked up by hand as long as you don t touch the tip Thi...

Страница 5: ...of the cantilever onto a 4 quadrant photodetector that measures any small displacements of the cantilever A piezoelectric Z scanner controls the height of the probe while a piezoelectric XY stage ras...

Страница 6: ...mage due to surface charging 3 0 Setup 3 1 Where to set it up The nGauge AFM is small simple and portable and can be set up just about anywhere However for best results it is recommended that you adhe...

Страница 7: ...ng it up near sources of electrical noise such as server rooms or rooms operating high power equipment 3 2 How to set it up Setup of the nGauge AFM is very simple 1 Remove the components from the carr...

Страница 8: ...nGauge AFM User Manual 8 3 Connect the USB cable to a computer and to the USB connector in the stage 4 Completely unscrew and remove the thumb nut from the stage...

Страница 9: ...between the tip and platform 7 Replace the thumb nut and firmly tighten it 3 3 Installing the Software The software required to operate the AFM can be found on the USB stick provided or downloaded fro...

Страница 10: ...ftware Overview Launch the software by double clicking on the nGauge program on your desktop or searching for nGauge in the start menu If the AFM is powered up and connected to the computer with the U...

Страница 11: ...the platform is low enough Position the sample so that the area to be scanned is directly under the tip Ensure the sample is stable on the sample platform and that it won t tilt roll flutter drift or...

Страница 12: ...ls above each point of the scan and ultimately determines the scan speed A smaller number in this field increases the scan speed but if the scan is too fast the resulting image may be blurry or distor...

Страница 13: ...s more force A lower setpoint can often give better results but is also more likely to damage the tip or sample and when the setpoint gets too low the imaging performance may suffer A PI controller is...

Страница 14: ...want to extract some quantitative data about the sample like the height or width of a specific feature or the RMS roughness of the surface These measurements are best acquired using Gwyddion a powerf...

Страница 15: ...ay a 3D view of the image You can rotate or scale this 3D image change the colors and save it Once you are satisfied with the post processed image you can save it as JPG PNG etc using the Save As opti...

Страница 16: ...still flat and noisy the AFM chip may be damaged Click on the Menu on the top right of the screen and scroll down and click Read All If any of the values under Resistances is greater than 1000 the AFM...

Страница 17: ...The main controls that are visible in each screen should be sufficient for normal imaging conditions However each tab in the navigation contains advanced controls that can extend the functionality or...

Страница 18: ...e line is being scanned repeatedly Leveling Direction To correct for the 15 degree tilt of the AFM chip relative to the sample a hardware leveling feature is applied Leveling Direction switches betwee...

Страница 19: ...of the Coarse and Fine Z the amplitude of the AC driving signal on the Z actuator and the amount of leveling that is added to each scan PGA values can range from 0 140 ADCs Analog to Digital Converte...

Страница 20: ...Port release the USB connection Choose Firmware currently Windows only Opens a window that allows you to locate the firmware update Firmware update files are binary bin files and can be downloaded fro...

Отзывы: