nGauge AFM User Manual

 

18

 

 

7.2 Approach - Advanced Controls 

 

In  the  approach  tab,  click  on

  Show

 

Advanced  Controls

.  These  additional  controls  are  for  the 

controller, and the coarse Z positioning. 

Proportional 

The proportional gain coefficient of the PI controller. See section 4.3 Imaging 
for details on tuning the controller. 

Integral 

The integral gain coefficient of the PI controller. See section 4.3 Imaging for 
details on tuning the controller. 

Setpoint 

The tip oscillation amplitude that the controller will try to maintain. Increasing 
this value can be thought of as applying less force between the tip and sample. 
Decreasing this value can be thought of as applying more force between the tip 
and sample. 

Enable/Disable 

Manually turns the controller on or off.  

Z Coarse 

Controls the coarse Z tip position. Increasing this value retracts the tip away 
from the sample, decreasing this value approaches the tip towards the sample. 
This controls the AFMc chip's integrated Z scanner and should not be confused 
with the stage controls that raise or lower the sample platform. 

 
 

7.3 Scan - Advanced Controls 

 

In the scan tab, click on 

Show Advanced Controls

.  

Scan Mode 

Switch between the normal "Area" mode which scans a square 2D area, and 
"Line" mode which scans a long 1D line. The "Line" mode results will still be 
displayed as a 2D image, even though only a single line is being scanned 
repeatedly. 

Leveling Direction 

To correct for the 15 degree tilt of the AFM chip relative to the sample, a 
hardware leveling feature is applied. Leveling Direction switches between 
adding or subtracting the 15 degree tilt and is device-specific. Different types of 
AFM chip may require different Leveling Directions. 

Leveling 

If the raw data of an image still contains significant tilt in the Y direction, This 
value can be increased or decreased to compensate.  

Z Coarse 

The Z actuator is split into two controls: Coarse Z and Fine Z. Fine Z has a 
smaller range and is the channel that the PI controller uses to follow the 
sample topology during a scan. Coarse Z has a large range and is simply used to 
coarsely position the Z actuator so that that the entire sample topography is in 
the range of the Fine Z. 

Proportional 

The proportional coefficient of the PI controller. Increasing this value increases 
the response time of the controller, but increasing it too much can cause 
instability in the Z actuator. 

Integral 

The integral coefficient of the PI controller. Increasing this value increases the 
response time of the controller, but increasing it too much can cause instability 
in the Z actuator. 

Setpoint 

The tip oscillation amplitude that the controller will try to maintain. A lower 
setpoint is analogous to applying a larger force between the tip and sample. 

Enable/Disable 

Turns on or off the PI controller that maintains a constant tip oscillation 
amplitude equal to the setpoint by controlling the Z actuator (Fine Z). 

Содержание nGauge AFM

Страница 1: ...User Manual Version 1 1 Atomic Force Microscope...

Страница 2: ...3 1 Where to set it up 6 3 2 How to set it up 7 3 3 Installing the Software 9 4 0 Basic Operation 10 4 1 Software Overview 10 4 2 Preparation and Approach 11 4 3 Imaging 12 4 4 Analyzing Data 14 5 0...

Страница 3: ...ollowing warnings and precautions should be observed to ensure the safety of the user and the instrument Use only the power supply provided The driving electronics are designed to accept 7 5Vdc and an...

Страница 4: ...ng DOWN in the storage box Use tweezers to pick them up The AFM chip can be placed on a flat surface with the tip facing up The AFM chip can be picked up by hand as long as you don t touch the tip Thi...

Страница 5: ...of the cantilever onto a 4 quadrant photodetector that measures any small displacements of the cantilever A piezoelectric Z scanner controls the height of the probe while a piezoelectric XY stage ras...

Страница 6: ...mage due to surface charging 3 0 Setup 3 1 Where to set it up The nGauge AFM is small simple and portable and can be set up just about anywhere However for best results it is recommended that you adhe...

Страница 7: ...ng it up near sources of electrical noise such as server rooms or rooms operating high power equipment 3 2 How to set it up Setup of the nGauge AFM is very simple 1 Remove the components from the carr...

Страница 8: ...nGauge AFM User Manual 8 3 Connect the USB cable to a computer and to the USB connector in the stage 4 Completely unscrew and remove the thumb nut from the stage...

Страница 9: ...between the tip and platform 7 Replace the thumb nut and firmly tighten it 3 3 Installing the Software The software required to operate the AFM can be found on the USB stick provided or downloaded fro...

Страница 10: ...ftware Overview Launch the software by double clicking on the nGauge program on your desktop or searching for nGauge in the start menu If the AFM is powered up and connected to the computer with the U...

Страница 11: ...the platform is low enough Position the sample so that the area to be scanned is directly under the tip Ensure the sample is stable on the sample platform and that it won t tilt roll flutter drift or...

Страница 12: ...ls above each point of the scan and ultimately determines the scan speed A smaller number in this field increases the scan speed but if the scan is too fast the resulting image may be blurry or distor...

Страница 13: ...s more force A lower setpoint can often give better results but is also more likely to damage the tip or sample and when the setpoint gets too low the imaging performance may suffer A PI controller is...

Страница 14: ...want to extract some quantitative data about the sample like the height or width of a specific feature or the RMS roughness of the surface These measurements are best acquired using Gwyddion a powerf...

Страница 15: ...ay a 3D view of the image You can rotate or scale this 3D image change the colors and save it Once you are satisfied with the post processed image you can save it as JPG PNG etc using the Save As opti...

Страница 16: ...still flat and noisy the AFM chip may be damaged Click on the Menu on the top right of the screen and scroll down and click Read All If any of the values under Resistances is greater than 1000 the AFM...

Страница 17: ...The main controls that are visible in each screen should be sufficient for normal imaging conditions However each tab in the navigation contains advanced controls that can extend the functionality or...

Страница 18: ...e line is being scanned repeatedly Leveling Direction To correct for the 15 degree tilt of the AFM chip relative to the sample a hardware leveling feature is applied Leveling Direction switches betwee...

Страница 19: ...of the Coarse and Fine Z the amplitude of the AC driving signal on the Z actuator and the amount of leveling that is added to each scan PGA values can range from 0 140 ADCs Analog to Digital Converte...

Страница 20: ...Port release the USB connection Choose Firmware currently Windows only Opens a window that allows you to locate the firmware update Firmware update files are binary bin files and can be downloaded fro...

Отзывы: