background image

 

 
 

 

nGauge AFM User Manual

 

17

 

 

used. It is also possible that the sample topology is not centered around 
the middle of the Fine Z range, which could result in image clipping. The 
Approach tab displays a graph with the Fine Z position. If it appears that 
most of the image data is at the lower or higher end of the Fine Z range, 
you can click 

Show Advanced Controls

 and increase or decrease the 

Coarse Z to center the data. In some cases, you may even need to 

Disengage

 and then re-

Approach. 

 

PID Controller Tuning 

The PI controller measures the tip oscillation amplitude and subtracts this 
value from the setpoint to give an error term. The error term is multiplied 
by the Proportional (P) coefficient and the Integral (I) coefficient to 
generate a response in the Fine Z. If the P and I coefficients are too low, 
the controller response will be slow and the tip will not track the sample 
topology very well. If the P and I coefficients are too high, the controller 
may oscillate and the image will look like noise. One method of tuning the 
P and I values is to set them both to zero and increase the P until it begins 
to oscillate, then decrease P by a bit. Then begin to increase the I until it 
begin to oscillate and decrease I by a bit. For more detail, refer to 

https://en.wikipedia.org/wiki/Ziegler%E2%80%93Nichols_method

 

7.0 Advanced Controls 

 

The  main  controls  that  are  visible  in  each  screen  should  be  sufficient  for  normal  imaging 

conditions.  However,  each  tab  in  the  navigation  contains  advanced  controls  that  can  extend  the 
functionality or control of the scan. In addition, there is a side menu available by clicking the button on 
the top right of the screen that will enable access to all of the controls available. 

7.1 Frequency Sweep - Advanced Controls 

 

In the frequency sweep tab, click on 

Show Advanced Controls

. These additional controls are for 

running a manual frequency sweep and controlling the driving frequency and tip oscillation amplitude. 
When  the  advanced  controls  are  displayed,  the  drive  frequency  can  also  be  set  by  clicking the sweep 
graph at the desired frequency. 
  

Start Frequency 

The frequency at which you would like to start a manual frequency sweep 

Stop Frequency 

The frequency at which you would like to stop a manual frequency sweep 

Step Size 

The step size, in hertz, that the manual frequency sweep will increment. Step Size 
determines the number of data points in the results.  

Manual Sweep 

Manual Sweep will perform a frequency sweep from Start Frequency to Stop 
Frequency, incrementing the frequency by Step Size. The results will be displayed 
in the main Frequency Sweep plot.  

Clear 

Clears all the results from the main Frequency Sweep plot. 

Amplitude (%) 

Controls the amplitude of the driving signal applied to the Z actuator. This can be 
thought of as the tip oscillation amplitude. Increasing this value will increase the 
tip oscillation amplitude.  

 

Содержание nGauge AFM

Страница 1: ...User Manual Version 1 1 Atomic Force Microscope...

Страница 2: ...3 1 Where to set it up 6 3 2 How to set it up 7 3 3 Installing the Software 9 4 0 Basic Operation 10 4 1 Software Overview 10 4 2 Preparation and Approach 11 4 3 Imaging 12 4 4 Analyzing Data 14 5 0...

Страница 3: ...ollowing warnings and precautions should be observed to ensure the safety of the user and the instrument Use only the power supply provided The driving electronics are designed to accept 7 5Vdc and an...

Страница 4: ...ng DOWN in the storage box Use tweezers to pick them up The AFM chip can be placed on a flat surface with the tip facing up The AFM chip can be picked up by hand as long as you don t touch the tip Thi...

Страница 5: ...of the cantilever onto a 4 quadrant photodetector that measures any small displacements of the cantilever A piezoelectric Z scanner controls the height of the probe while a piezoelectric XY stage ras...

Страница 6: ...mage due to surface charging 3 0 Setup 3 1 Where to set it up The nGauge AFM is small simple and portable and can be set up just about anywhere However for best results it is recommended that you adhe...

Страница 7: ...ng it up near sources of electrical noise such as server rooms or rooms operating high power equipment 3 2 How to set it up Setup of the nGauge AFM is very simple 1 Remove the components from the carr...

Страница 8: ...nGauge AFM User Manual 8 3 Connect the USB cable to a computer and to the USB connector in the stage 4 Completely unscrew and remove the thumb nut from the stage...

Страница 9: ...between the tip and platform 7 Replace the thumb nut and firmly tighten it 3 3 Installing the Software The software required to operate the AFM can be found on the USB stick provided or downloaded fro...

Страница 10: ...ftware Overview Launch the software by double clicking on the nGauge program on your desktop or searching for nGauge in the start menu If the AFM is powered up and connected to the computer with the U...

Страница 11: ...the platform is low enough Position the sample so that the area to be scanned is directly under the tip Ensure the sample is stable on the sample platform and that it won t tilt roll flutter drift or...

Страница 12: ...ls above each point of the scan and ultimately determines the scan speed A smaller number in this field increases the scan speed but if the scan is too fast the resulting image may be blurry or distor...

Страница 13: ...s more force A lower setpoint can often give better results but is also more likely to damage the tip or sample and when the setpoint gets too low the imaging performance may suffer A PI controller is...

Страница 14: ...want to extract some quantitative data about the sample like the height or width of a specific feature or the RMS roughness of the surface These measurements are best acquired using Gwyddion a powerf...

Страница 15: ...ay a 3D view of the image You can rotate or scale this 3D image change the colors and save it Once you are satisfied with the post processed image you can save it as JPG PNG etc using the Save As opti...

Страница 16: ...still flat and noisy the AFM chip may be damaged Click on the Menu on the top right of the screen and scroll down and click Read All If any of the values under Resistances is greater than 1000 the AFM...

Страница 17: ...The main controls that are visible in each screen should be sufficient for normal imaging conditions However each tab in the navigation contains advanced controls that can extend the functionality or...

Страница 18: ...e line is being scanned repeatedly Leveling Direction To correct for the 15 degree tilt of the AFM chip relative to the sample a hardware leveling feature is applied Leveling Direction switches betwee...

Страница 19: ...of the Coarse and Fine Z the amplitude of the AC driving signal on the Z actuator and the amount of leveling that is added to each scan PGA values can range from 0 140 ADCs Analog to Digital Converte...

Страница 20: ...Port release the USB connection Choose Firmware currently Windows only Opens a window that allows you to locate the firmware update Firmware update files are binary bin files and can be downloaded fro...

Отзывы: