![ICS 8099 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/ics/8099/8099_instruction-manual_3540995044.webp)
3-8
3
3.3.7 Saving the Enable and Transition Register Values
When the PSC flag is set, the Enable and Transition Register values are cleared
at power turn-on. The registers can
only
be saved and recalled at power turn-on
by disabling the PSC flag. The *SAV command does not save these registers.
Use the
*PSC 0
command to disable the PSC flag and save the current Enable
and Transition register values as shown in the example. e.g.
STAT:OPER:ENAB 1
'enables Status A bit
STAT:OPER:NTR 1
'enables negative transition
*PSC 0; ESE 192; SRE 32
'saves Power-on and EDR
bits and current registers
values as the new power on
settings.
The enable and transition register setting commands must be on the same line
or set prior to the
*PSC 0
command to be saved. A later
*PSC 1
command
sets the PSC flag which will cause the registers to be cleared at the next power
turn-on.
3.3.8 488.2 Differences from 488.1 Devices
The IEEE 488.1 Device Clear command
does not
reset the digital outputs as
would be expected of a 488.1 device. To reset the digital outputs, use the
*RST
(Reset) or
*RCL 0
command.