2-8
2
7.
Your new settings have been saved in the board’s flash memory. You
have to reboot the unit or power cycle it for the changes to take affect.
Press the ‘Reboot’ button to reboot the unit now. The ‘Return to the
Configuration Page’ button only works if you did not change the unit’s
IP address.
2.5.2 VXI-11 Configuration Utility Method
This method uses ICS’s VXI-11 Configuration Utility program, VXI11_con-
fig, to view and change the current network and serial configuration settings.
VXI11_config runs in WIN32 PC (Windows 98, Me, 2K, XP and 2003 operating
systems. VXI11_config is included on the Support CD-ROM supplied with the
unit. Select the Install VXI-11 Support option on the Support CD’s Main Page
to install VXI11_config and VXI11_kybd on your computer. VXI11_config
can be run from Window’s Start menu by pointing to Programs and then to
ICS_Electronics. Select VXI11_config from the submenu.
1.
Use the Crossover Cable to connect the 8099 directly to the computer
running the browser as shown in Figure 2-8.
PC with a Windows
Operating System
Ethernet
Crossover
Cable
8099 Ethernet to M
odbusl
8099 Modbus Interface
Figure 2-8 8099 Connected to computer with a Crossover Cable
Alternately use a standard Ethernet Cable to connect the card to the
same hub or switch that the computer running the browser is con-
nected to as shown in Figure 2-1. Temporarily disconnect any local
network connection to avoid network conflicts until the module is
configured.
2.
Apply power to the 8099 and verify that it passes its selftest routine
and that the PWR, RDY and LAN LEDs are on.
3.
Check your PC’s network settings to be sure its IP address is in the
192.168.0.xxx range so it can communicate with the card’s default IP
address. To check, right-click on My Network Places and click on
Properties. Right-click on Local Area Connection and click Properties.