INTRODUCTION

 

INTRODUCTION

 

 

 

LCD Display Signal

 

 
 

CAUTION 

 

 

This device is intended for adult use only.

 

This device is intended for no-invasive measuring and monitoring of arterial blood pressure.

 

It is not intended for use on extremities other than the arm or for functions other than obtaining a

 

blood pressure measurement.

 

Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood

 

pressure.Do not begin or end medical treatment without asking a physician for treatment advice.

 

If you are taking medication,consult your physician to determine the most appropriate time to

 

measure your blood pressure. Never change a prescribed medication without consulting your

 

Physician.

 

When the device was used to measure patients who have common arrhythmias such as atrial or

 

ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please

 

consult your physician about the result.

 

If the cuff pressure exceeds 40 kPa (300 mmHg), the unit will automatically deflate. Should the 
cuff not deflate when pressures exceeds 40 kPa (300 mmHg), detach the cuff from the arm and 
press the corresponding user button to stop inflation.

 

The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable

 

anesthetic mixture with air of with oxygen or nitrous oxide.

 

The operator shall not touch output of batteries /adapter and the patient simultaneously.

 

To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated

 

interference signal or electrical fast transient/burst signal.

 

The user must check that the equipment functions safely and see that it is in proper working

 

condition before being used.

 

This device is contraindicated for any female who may be suspected of, or is pregnant. Besides

 

providing inaccurate readings, the effects of this device on the fetus are unknown.

 

Manufacturer will make available on request circuit diagrams, component parts list etc.

 

This unit is not suitable for continuous monitoring during medical emergencies or operations.

 

Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and even purple due to

 

a lack of blood.

 

Please use the device under the environment which was provided in the user manual. Otherwise,

 

the performance and lifetime of the device will be impacted and reduced.

 

During use, the patient will be in contact with the cuff. The materials of the cuff have been tested

 

and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not

 

cause any potential sensization or irritation reaction.

 

Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.

 

Otherwise, it may cause damage to the unit or danger to the user/patients.

 

The device doesn’t need to be calibrated within the two years of reliable service.

 

Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the

 

local guidelines.

 

If you have any problems with this device, such as setting up, maintaining or using, please

 

contact the SERVICE PERSONNEL of iChoice. Don’t open or repair the device by yourself.

 

Please report to iChoice if any unexpected operation or events occur.

 

Please use the soft cloth to clean the whole unit. Don’t use any abrasive or volatile cleaners.

 

 
 

 

5

 

SYMB
OL

 

DESCRIPTION

 

EXPLANATION

 

 

Systolic Blood Pressure

 

High blood pressure

 

 

 

Diastolic Blood Pressure

 

Low blood pressure

 

 

Pulse per minute

 

Beats per minute, BPM

 

 

 

User 1

 

Start measurement save and transmit 
the

 

measuring result  for user 1

 

 

 

User 2

 

Start measurement, save and transmit 
the

 

measuring result for user 2

 

 

Heartbeat

 

Heartbeat detection during the measurement

 

 

Data Pending to Transmit

 

Measurement data stored in the 
equipment

 

 

Data Transmitting

 

Data is transmitting

 

 

Average Value

 

Average value of last three measurements

 

 

Shocking reminder

 

Shocking will result in inaccurate

 

 

Low Battery

 

Low battery and please replace the batteries

 

 

Unit

 

Measurement unit of blood pressure

 

(1mmHg=0.133kPa) 

(1kPa=7.5mmHg)

 

 

Current Time

 

Month/Day (Hour : Minute)

 

 

Deflating

 

Exhaust the air in the cuff

 

 

Data Enquiry Mode

 

Recall the records

 

 

Irregular Heartbeat

 

Irregular Heartbeat Detection

 

 

Memory

 

Display  the  serial  number  of 
the

 

measurement

 

 

Содержание LS805-B

Страница 1: ...Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Arm Type Contains FCC I...

Страница 2: ...the Records Delete the Records Data Transmission INFORMATION FOR USER 15 Tips for measurement Maintenance ABOUT BLOOD PRESSURE 17 What are systolic pressure and diastolic pressure What is the standar...

Страница 3: ...acemaker or defibrillator 2 The device is not intended to be a diagnostic device Contract your physician if hypertensive values are indicated Measurement Principle This product uses the Oscillometric...

Страница 4: ...continuous monitoring during medical emergencies or operations Otherwise the patient s arm and fingers will become anaesthetic swollen and even purple due to a lack of blood Please use the device und...

Страница 5: ...ttery door List 1 Blood Pressure Monitor 3 4 AAA batteries 6 2 Cuff Type BF applied part 22cm 32cm Please use iChoice authorized cuff The size of the actual cuff please refer to the label on the attac...

Страница 6: ...ss Query button to change the numeral Each press 4 Repeat steps 2 and 3 to confirm the MINUTE Then the monitor diverts to MONTH and DAY setting MONTH DAY 5 Repeat steps 2 and 3 to confirm MONTH and DA...

Страница 7: ...cuff s air pipe into the interface located on the right side of the monitor 2 As pictured on the right wear the cuff on your upper arm 3 Tighten the cuff up Make sure the cuff is fixed 2 to 3 centime...

Страница 8: ...ther user button to switch to display another user s measurement data 4 When User 2 has no record the LCD will display just like the picture on the right 5 Press User 1 User 2 button to turn off the m...

Страница 9: ...device 3 If the data transmission fails the symbol will remain The pending measurement data will be transmitted to your mobile device when next measurement is complete Within 1 hour after dinner or dr...

Страница 10: ...uideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mm Hg Diastolic Blood Pressure 85 mm Hg Irregular Heartbeat Detector An irregular heartbeat is detected when a heartbeat rhythm varies...

Страница 11: ...breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down It is ok for both arms but there will be some different results for different people We...

Страница 12: ...ad Haidian District 100143 Beijing P R China Complied Standards List 21 Risk Management ISO 14971 2007 Labeling EN 980 2008 User Manual EN 1041 2008 Generl Requirements for Safety IEC 60601 1 2005 A1...

Страница 13: ...electronic equipment may be affected RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manu...

Страница 14: ...d below The customer or the user of the device should assure that it is used in such an environment IMMUNITY test IEC 60601 TEST LEVEL Complia nce level Electromagnetic environment guidance Conducted...

Отзывы: