Table 4 Guidance and MANUFACTURER’s declaration – electromagnetic IMMUNITY –

 

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

 

 

Table 6 Recommended separation distances between portable and mobile

 

RF communications equipment and the ME EQUIPMENT or ME SYSTEM –

 

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

 

 

Recommended separation distances

 

between  portable and mobile RF communications equipment and the device.

 

The device is intended for use in an electromagnetic environment in which

    

radiated RF disturbances are controlled. The customer or the user of the device

 

can help prevent electromagnetic interference by maintaining a minimum distance

 

between portable and mobile RF communications equipment (transmitters) and the 
device as recommended below, according to the maximum output power of the

 

communications equipment.

 

Rated maximum output

 

power of transmitter

 

(W)

 

Separation distance according to frequency of transmitter (m)

 

150 kHz to 80 MHz

 

d = 

1.167  

P

 

80 MHz to 800 MHz

 

d = 

1.167  

P

 

800 MHz to 2.5 GHz

 

d = 

2.333   

P

 

0.01

 

0.117

 

0.117

 

0.233

 

0.1

 

0.369

 

0.369

 

0.738

 

1

 

1.167

 

1.167

 

2.333

 

10

 

3.690

 

3.690

 

7.378

 

100

 

11.67

 

11.67

 

23.33

 

For transmitters rated at a maximum output power not listed above, the recommended

 

separation distance 

in metres (m) can be estimated using the equation applicable to 

the frequency of the transmitter, where 

is the maximum output power rating of the

 

transmitter in watts (W) according to the transmitter manufacturer.

 

NOTE 1    At 80MHz and 800MHz, the separation distance for the higher frequency

 

range applies.

 

NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation

 

is affected by absorption and reflection from structures, objects and people.

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity

 

The device is intended for use in the electromagnetic environment specified below.

 

The customer or the user of the device should assure that it is used in such an environment.

 

IMMUNITY test

 

IEC 60601

 

TEST 
LEVEL

 

Complia
nce

 

level

 

Electromagnetic 
environment -

 

guidance

 

 
 
 
 
 
 
 
 

Conducted RF

 

IEC 61000-4-6

 

 

Radiated RF

 

IEC 61000-4-3

 

 
 
 
 
 
 
 
 

3 Vrms

 

150 kHz to

 

80 MHz

 

 

3 V/m

 

80 MHz to

 

2.5 GHz

 

 
 
 
 
 
 
 
 

 

3 Vrms

 

 
 

3 V/m

 

Portable and mobile RF 
communications

 

equipment should 

be used no closer to

 

any part of the 

device, including

 

cables, than the recommended

 

separation distance calculated 
from the

 

equation applicable to the 

frequency of

 

the transmitter.

 

Recommended separation distance

 

d = 

1.167  

P

 

 

d  = 

1.167  

P  

80  MHz  to  800  MHz

 

 

d = 

2.333   

P  

800  MHz  to  2,5  GHz

 

where 

is the maximum output 

power

 

rating of the transmitter in 

watts (W) according to the 
transmitter manufacturer

 

and 

is the 

recommended separation

 

distance in 

metres (m).

 

Field strengths from fixed RF

 

transmitters, as determined by an

 

electromagnetic site survey, 

should be

 

less than the compliance 

level in each

 

frequency range.

b

 

Interference may occur in the 
vicinity of

 

equipment marked with 

the following

 

symbol:

 

NOTE 1

 

At 80 MHz and 800 MHz, the higher frequency range applies.

 

NOTE 2

 

 

These guidelines may not apply in all situations. Electromagnetic

  

propagation 

is affected by absorption and reflection from structures, objects

 

and people.

 

Field strengths from fixed transmitters, such as base stations for radio (cellular /

 

cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad- 
cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the

 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site

 

survey should be considered. If the measured field strength in the location in which  the

 

device is used exceeds the applicable RF compliance level above, the device should

 

be 

observed to verify normal operation. If abnormal performance is observed,

 

additional 

measures may be necessary, such as re-orienting or relocating the device.

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

 

 

EMC GUIDANCE 

EMC GUIDANCE 

Содержание LS805-B

Страница 1: ...Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Arm Type Contains FCC I...

Страница 2: ...the Records Delete the Records Data Transmission INFORMATION FOR USER 15 Tips for measurement Maintenance ABOUT BLOOD PRESSURE 17 What are systolic pressure and diastolic pressure What is the standar...

Страница 3: ...acemaker or defibrillator 2 The device is not intended to be a diagnostic device Contract your physician if hypertensive values are indicated Measurement Principle This product uses the Oscillometric...

Страница 4: ...continuous monitoring during medical emergencies or operations Otherwise the patient s arm and fingers will become anaesthetic swollen and even purple due to a lack of blood Please use the device und...

Страница 5: ...ttery door List 1 Blood Pressure Monitor 3 4 AAA batteries 6 2 Cuff Type BF applied part 22cm 32cm Please use iChoice authorized cuff The size of the actual cuff please refer to the label on the attac...

Страница 6: ...ss Query button to change the numeral Each press 4 Repeat steps 2 and 3 to confirm the MINUTE Then the monitor diverts to MONTH and DAY setting MONTH DAY 5 Repeat steps 2 and 3 to confirm MONTH and DA...

Страница 7: ...cuff s air pipe into the interface located on the right side of the monitor 2 As pictured on the right wear the cuff on your upper arm 3 Tighten the cuff up Make sure the cuff is fixed 2 to 3 centime...

Страница 8: ...ther user button to switch to display another user s measurement data 4 When User 2 has no record the LCD will display just like the picture on the right 5 Press User 1 User 2 button to turn off the m...

Страница 9: ...device 3 If the data transmission fails the symbol will remain The pending measurement data will be transmitted to your mobile device when next measurement is complete Within 1 hour after dinner or dr...

Страница 10: ...uideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mm Hg Diastolic Blood Pressure 85 mm Hg Irregular Heartbeat Detector An irregular heartbeat is detected when a heartbeat rhythm varies...

Страница 11: ...breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down It is ok for both arms but there will be some different results for different people We...

Страница 12: ...ad Haidian District 100143 Beijing P R China Complied Standards List 21 Risk Management ISO 14971 2007 Labeling EN 980 2008 User Manual EN 1041 2008 Generl Requirements for Safety IEC 60601 1 2005 A1...

Страница 13: ...electronic equipment may be affected RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manu...

Страница 14: ...d below The customer or the user of the device should assure that it is used in such an environment IMMUNITY test IEC 60601 TEST LEVEL Complia nce level Electromagnetic environment guidance Conducted...

Отзывы: