Tip 1:
Error Code 15 Potential Causes
Water level sensor (high-
level probe) is not detected
after 5 minutes.
Water:
•
Is the water supply to the machine open?
•
Is the water supply pressure below 20psi?
•
Does the water inlet solenoid have power to it from PCB?
•
Is the water inlet solenoid valve clogged with scale/debris?
•
Is there a water filter that’s overdue to be changed?
•
Does the water level sensor appear scaled up and in need of
cleaning?
•
Is water overflowing the water trough, indicating the water
level sensor is not sensing a full trough?
•
Is the trough full of water, but the water level sensor is
indicating that the machine is empty?
•
Is the dump valve stuck open?
IM-Series
Error Codes
45