MESSAGE

DESCRIPTION

ACTIONS

Battery charge is low.

Prepare to replace the batteries.

The batteries are empty.

Turn the tonometer OFF by pressing Select button.
Replace the batteries.

The probe did not move.

Change the probe. The probe may be twisted or otherwise inserted 
incorrectly. To clear error messages, press the Measure button, after which 
the measurement can be repeated.

The probe did not move properly for several times during 
the measurement sequence.

Remove and clean the probe base or replace it with new one as instructed in 
Replacing/cleaning the probe base. To clear error messages, press the Mea-
sure button, after which the measurement can be repeated.

The probe did not touch the eye.

Adjust correct measurement distance 4-8 mm (5/32-5/16 inch). 
The measurement was taken from too far away.

Too short measurement distance between the probe and 
the cornea.

Adjust correct measurement distance 4-8 mm (5/32- 5/16 inch).  
The measurement was taken from too close. To clear error messages, press 
the Measure button, after which the measurement can be repeated.

The probe did not move properly. The probe did not make 
clean contact with the cornea, because the probe hit an 
eyelid or eyelashes.

Ensure that the eye is open, measure again. To clear error messages, press the 
Measure button, after which the measurement can be repeated.

Internal error detected.

Turn the tonometer OFF by pressing Select button. Contact the seller to
arrange sending the device for service.

ERROR AND INFO MESSAGES

 

The following messages may appear on the display:

MENU FUNCTIONS

    

Scrolling between the Menu functions starts from the MEASURE display, press either of the Navigation buttons located

 

   

around the Select button.

 

   

Menu functions are MEASURE, SETTING, HISTORY, SOUND, LIGHT, LANGUAGE and INFO.

    

 

 

MEASURE – Access to measurement

 

   

 

 

Press the Select button to access.

 

   

 

 

If the probe is not loaded the LOAD display appears.

 

   

 

 

Tonometer is ready for measurement when Play-symbol display appears.

 

   

 

 

To exit, press the Select button. 

    

 

 

SETTING – Selecting specie

 

   

 

 

Previously selected specie symbol is displayed. To select another specie, 

 

   

 

 

press the Select button and scroll throught the settings by pressing either  

 

   

 

 

of the Navigation buttons.

 

   

 

 

To accept selection, press the Select button.    

 

 

    

 

 

HISTORY – Old measurements

 

   

 

 

Press the Select button to access.

 

   

 

 

Scroll through the old values by pressing either of the Navigation buttons.

 

   

 

 

Value colors green, yellow and red are related to Standard deviation (SD).

 

   

 

 

To exit, press the Select button.

    

 

 

SOUND – Setting of Tonometer buzzer

 

   

 

 

Green text and symbol is active setting.

 

   

 

 

Press the Select button to access.

 

   

 

 

Select the volume setting by pressing either of the Navigation buttons.

 

   

 

 

To accept selection, press the Select button.

    

 

 

LIGHT – Setting of Probe base light (positioning guide)

 

   

 

 

Green text and symbol is active setting.

 

   

 

 

Press the Select button to access.

 

   

 

 

Turn the light ON and OFF by pressing either of the Navigation buttons.

 

   

 

 

To accept selection, press the Select button.

    

 

 

LANGUAGE – Language setting

 

   

 

 

Green text is active setting.

 

   

 

 

Press the Select button to access.

 

   

 

 

Scroll through the language options by pressing either of the Navigation buttons.

 

   

 

 

To accept selection, press the Select button.

    

 

 

INFO – Device information

 

   

 

 

Press the Select button to access.

 

   

 

 

Serial number (SN) of the tonometer.

 

   

 

 

Software version (SW) of the tonometer.

 

   

 

 

To exit, press the Select button.

TURNING THE TONOMETER OFF

Press the Select button until the display shows the End-symbol.

If you do not use the tonometer, it will switch off automatically after 3 minutes.

Содержание TONOVET PLUS

Страница 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Страница 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Страница 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Страница 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Страница 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Страница 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Страница 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Страница 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Страница 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Отзывы: