Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

www.tonovet.com

10

www.tonovet.com

11

ACCESSORIES

TECHNICAL INFORMATION

Type: TV011
Dimensions: 24 - 29 mm (W) * 35 - 95 mm (H) * 215 mm (L)
Weight: 140 g (without batteries), 230 g (4 x AA batteries)
Power supply: 4 x AA non-rechargeable batteries, 1.5V alkaline LR6

Measurement range: 10 - 60 mmHg
Accuracy: ± 2.5 mmHg (10 - 30 mmHg) and ± 10% (>30 - 60 mmHg)
Repeatability (coefficient of variation): < 8 % 
Accuracy of display: 1 mmHg
Display unit: Millimeter of mercury (mmHg)
The serial number is on the inside of the battery compartment cover.
There are no electrical connections from the tonometer to the patient.
The device has BF-type electric shock protection.

Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa - 1,060 hPa 

Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa - 1,060 hPa

 

Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 hPa - 1,060 hPa

 

Mode of operation: continuous

SKU

PRODUCT DESCRIPTION

WEIGHT DIMENSIONS

104

Box of 100 probes

89 g

53 x 109 x 36 mm

7217

Probe base collar

4 g

18 x 18 mm

7218

Probe base collar, narrow

1 g

17 x 18 mm

540

Probe base

4 g

7 x 38 mm

543

Probe base cleaning container

3 g

20 x 56 mm

544

Probe base cover 

1 g

19 x 11 mm

559

Wrist strap with lock

4 g

10 x 10 x 270 mm

526

Aluminium case

800 g

240 x 280 x 72 mm

7169

Battery cover & screw

6 g

110 x 25 x 12 mm

565

Silicone grip - green

26 g

45 x 35 x 113 mm

624

IOP pad

38 g

50 x 53 x 16 mm

577

USB memory stick (including instruction manuals as PDF)

44 g

98 x 11 x 93 mm

645

Quick guide

19 g

210 x 148 mm

548

Screw driver

15 g

16 x 90 mm

MEASUREMENT FLOW CHART

Turn Tonometer ON by pressing Select or Measure button

Load Probe

Ready to measure (NOTE the specie setting)

Successful 
measurement
with none or 
insignificant 
deviation.

Tonometer OFF by pressing 
Select button > 3 seconds

Measure 6 times by pressing Measure button
(green color bar shows the progress) 

This is displayed if you pressed the Select button, pressing of 
Measure button would lead you directly to Load Probe 

Deviation is too great,  
repeat the measurement.
NOTE High IOP may cause high 
fluctuation, which may cause 
higher standard deviation.

Acceptable measurement 
with slightly greater than 
normal standard deviation.
Repeating the measurement 
is recommended if the IOP is 
higher than normal.

Содержание TONOVET PLUS

Страница 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Страница 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Страница 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Страница 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Страница 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Страница 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Страница 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Страница 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Страница 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Отзывы: