GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY

Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic environment specified below. 

The customer or the user of the Icare TONOVET Plus tonometer (TV011) should assure that it is such an environment.

Immunity test

IEC 60601 Test level

Compliance level

Electromagnetic environment-Guidance

Radiated RF IEC 61000-
4-3

Conducted RF
IEC 61000-4-6

3 V/m
80MHz to 2,5 GHz 

3Vrms
150 kHz to 80 MHz

3V/m

NOT APPLICABLE

Portable and mobile RF communications equipment should be used no closer 
to any part of the Icare TONOVET Plus tonometer (TV011), including cables, 
than the recommended separation distance calculated from the equation 
applicable to the frequency of the transmitter.

Recommended separation distance

d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2 5 GHz

Where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer and d is the recommended 
separation distance in metres (m). 

Field strengths from fixed RF transmitters, as determined by an 
electromagnetic site survey 

a

 should be less than the compliance level  

in each frequency range 

b

.

Interference may occur in the vicinity of equipment  
marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM 

radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, 
an electromagnetic site survey should be considered. If the measured field strength in the location in which the Icare TONOVET Plus tonometer (TV011) is used 
exceeds the applicable RF compliance level above, the Icare TONOVET Plus tonometer (TV011) should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Icare TONOVET Plus tonometer (TV011).  

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND 

ICARE TONOVET PLUS TONOMETER

Icare TONOVET Plus tonometer (TV011) is intended for use in an electromagnetic environment in which radiated RF-disturbances are controlled. The customer 
or the user of the Icare TONOVET Plus tonometer (TV011) can help prevent electromagnetic interference by maintaining a minimum distance between portable 
and mobile RF communications equipment (transmitters) and Icare TONOVET Plus tonometer (TV011) as recommended below, according to the maximum 
output power of the communications equipment.

Rated maximum output power of 
transmitter

(W)

Separation distance according to frequency of transmitter, m

150 kHz to 80 MHz
d= 1,2 √P

80 MHz to 800 MHz
d= 1,2 √P

800 MHz to 2,5 GHz 
d= 2,3 √P

0,01

NOT APPLICABLE

0.12

0.23

0,1

NOT APPLICABLE

0.38

0.73

1

NOT APPLICABLE

1.2

2.3

10

NOT APPLICABLE

3.8

7.2

100

NOT APPLICABLE

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the 
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the 
transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

TV011 is class B equipment and needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided 
in Instruction for use manual.

GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC EMISSIONS

Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic environment specified below. 

The user of the Icare TONOVET Plus tonometer (TV011) should assure that it is used in such an environment.

RF emissions CISPR 11

Group 1

Icare TONOVET Plus tonometer (TV011) is battery operated and use RF energy only for its internal 
function. Therefore, its RF emissions are low and are not likely to cause any interference in nearby 
equipment.

RF emissions CISPR 11

Class B

Icare TONOVET Plus tonometer (TV011) is suitable for use in all establishments, including domestic 
establishments and those directly connected to public low-voltage power supply network that 
supplies buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

NOT APPLICABLE

Voltage fluctuations flickering 
emissions
IEC 61000-3-3

NOT APPLICABLE

GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC IMMUNITY

Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic enviroment specified below. 

The customers or users of Icare TONOVET Plus tonometer (TV011) should assurre that it is used in such enviroment.

Immunity test

IEC 60601 Test level

Compliance level

Electromagnetic environment-Guidance

Electrostatic discharge (ESD)  
IEC 61000-4-2

± 6 kV contact
±8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with 
synthetic material, the relative humidity should be at least 30 %

Electrical fast Transients/burst  
IEC 61000-4-4

±2 kV for power supply lines
±1 kV for input/output lines

NOT APPLICABLE

NOT APPLICABLE

Surge
IEC 61000-4-5

±1 kV for line(s) to line(s)
±2 kV for line(s) to earth

NOT APPLICABLE

NOT APPLICABLE

Voltage dips, short interruption 
and voltage variations on power 
supply lines
IEC 61000-4-11

<5 % UT (>95 % dip in UT) 
for 0,5 cycle
40 % UT (60 % dip in UT) 
for 5 cycles
70 % UT (30 % dip in UT) 
for 25 cycles
<5 % UT (>95 % dip in UT) 
for 5 s

NOT APPLICABLE

NOT APPLICABLE

Power frequency (50/60 Hz) 
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic of a 
typical location in a typical commercial or hospital environment.

Содержание TONOVET PLUS

Страница 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Страница 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Страница 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Страница 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Страница 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Страница 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Страница 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Страница 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Страница 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Отзывы: