manualshive.com logo in svg
background image

Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

www.tonovet.com

6

www.tonovet.com

7

STEP 5. 

Following the performance of the entire measurement, a new measurement series can be begun by pressing the Measure button. The tonometer will 

then reactivate the probe and be ready for the next measurement series with the Play symbol on the display. The measurement sequence can be aborted by 
pressing the Select button. Place the probe base plug back to cover the probe base when the tonometer is not in use.

  

STEP 3.

 You may perform the measurement in single or series mode. 

   

Single mode: Press the Measure button lightly to perform the measurement, taking care not to shake the tonometer. The tip of the probe  

   

should make contact with the central cornea. Six measurements should be made consecutively, green segments will be lit after every 

   

successful measurement. After each successful measurement, you will hear a short beep. 

   

Series mode: Keep the Measure button down to obtain the sequence of six measurements, green segments will be lit after every successful  

   

measurement.

To obtain the final reading, six measurements are required. The measurement values displayed before the final result are average values for all previous 
measurements (1.-5.). Single measurement values are not shown.

   

If there is an erroneous measurement, the tonometer will beep twice and display an error message. Press the Measure button to clear the  

   

error message. If several erroneous measurements appear, see error messages segment for additional information.

 
 
 
 
 

STEP 4.

 Once the six measurements have been performed, you will hear a long beep. The final IOP will be shown on the display rounded by green (succesfull) or yellow

(some variation) segments. If variation is too big, red segments will be displayed. Yellow Repeat symbol will be dispayed in case of external distubance (eg. EMC) during 
the measurement.

   

 

 

 

The displayed result is an average of four measurements as the highest and the lowest  

   

 

 

 

reading are discarded before the average calculation.

   

 

 

 

The colors green, yellow and red indication are related to the standard deviation (SD) of

    

 

 

 

the probe’s motion parameters of the four remaining measurements. When red 

   

 

 

 

segments are displayed; the measurement should be repeated.

  

 

   

NOTE High IOP may cause high fluctuation, which may cause higher standard deviation.

  

MEASUREMENT

 NOTE
If you do not use the tonometer, it will switch off automatically after 3 minutes.

 

NOTE

No anesthetic is required when performing measurements. 

 

NOTE

Avoid excessive restraining, as it may alter the IOP. Patient’s head should be held as lightly as possible; be careful not to put pressure on the neck or the eye ball. 
If a collar is worn; make sure it is not too tight or remove it for the measurement.

STEP 1.

 Bring the 

tonometer near the 
patient’s eye.

*Badouin C, Gastaud P. Influence of topical anesthesia on tonometeric values of intraocular pressure. Ophthalmologica 1994;208:309-313.

Correct alignment of the tonometer and green probe base light indication.

Incorrect alignment of the tonometer and red probe base light indication.

Correct head and eye position.

Incorrect head and eye position.

STEP 2.

 The probe should be in a horizontal position. Keep the probe horizontal and pointing perpendicularly to the center of the cornea. The distance from the tip 

of the probe to the patient’s cornea (see picture) should be 4-8 mm (5/32 - 5/16 inch).

If probe base light indication is set OFF 
green arrows on display will indicate the 
correct alignment of the tonometer.

If probe base light indication is set OFF red 
arrows on display will indicate the incorrect 
alignment of the tonometer.

Содержание TONOVET PLUS

Страница 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Страница 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Страница 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Страница 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Страница 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Страница 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Страница 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Страница 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Страница 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Отзывы: