background image

ENGLISH

Icare ic200 tonometer manual

www.icaretonometer.com

4

Keep the tonometer out of reach of children. The probe base, battery compartment cover, screws, collar and probes are small objects and may be accidentally 

swallowed.

Do not use the device if it appears to be damaged or malfunctioning. The device must be delivered to service for repair.

Do not use the device near inflammable substances, including inflammable anesthetic agents.

Prior to each new patient to be measured, check that a new disposable probe from an intact package is being used. After inserting the probe in the probe base, 

visually inspect the probe to ensure that the small plastic round tip is visible at the front. Do not use a probe without the plastic tip.

The tonometer conforms to EMC requirements (IEC 60101-1-2), performance of the tonometer may be affected if it is used near to (<1 m) another electrical 

device, such as a cellular phone, emitting high-intensity electromagnetic radiation. The tonometer’s own electromagnetic emissions are well below the maximum 

levels permitted by the relevant standards. Nevertheless, the tonometer may cause interference in the operation of highly sensitive devices in the immediate 

vicinity.

If the device is not intended to be used for a long period of time, it is recommended to remove the batteries from the battery compartment. Removing the 

batteries will not affect the subsequent functioning of the tonometer.

Used probes cannot be recycled. Dispose of used probes properly (e.g. in containers for disposable needles or in a bin for metal waste).

Batteries, packaging materials and probe bases must be disposed of according to local regulations.

To avoid dropping of the device and to ensure the safe handling of the device, always use the wrist strap to keep the device attached

to the wrist when in use.

Use only the types of battery specified in the Technical Information section of this instruction manual.

The measurement method of the Icare ic200 tonometer is based on magnetic induction and therefore an external magnetic field in line with the probe may 

prevent the measurement. In such case the tonometer will continuously ask to repeat the measurement. Situation can be solved either by removing the source 

of interference from the vicinity of the device or by performing the measurement in different location with no such interference.

Changes to the IT network could introduce new risks requiring additional analysis by the responsible organization. The changes include:

•  changes in the IT-network configuration

•  connection of additional items to the IT-network

•  disconnecting items from the IT-network

•  update or upgrade of equipment connected to the IT-network

All manuals and user guides at all-guides.com

Содержание ic200

Страница 1: ...INSTRUCTION MANUAL ENGLISH All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 2: ...OBE 7 PROBE BASE INDICATOR LIGHT 7 MEASUREMENT 8 USER INTERFACE FUNCTIONS 11 BLUETOOTH 12 PRINTER 12 EXPORT 13 ERROR AND INFO MESSAGES 14 MEASUREMENT FLOW CHART 15 ACCESSORIES 16 TECHNICAL INFORMATION...

Страница 3: ...nt in both eyes but in case either eye is inflamed or infected the healthy eye should be measured first Use probes taken only from the intact original packaging Re use of a probe could result in incor...

Страница 4: ...f time it is recommended to remove the batteries from the battery compartment Removing the batteries will not affect the subsequent functioning of the tonometer Used probes cannot be recycled Dispose...

Страница 5: ...ters covering the motion of the probe including deceleration and rebound time Using a proprietary algorithm the device is able to calculate the eye s IOP A displayed IOP reading is derived from the re...

Страница 6: ...w the battery compartment locking screw with the screwdriver supplied Remove the battery compartment cover Insert a new set of four AA 1 5 V batteries LR6 Insert the batteries according to the figure...

Страница 7: ...f the tonometer and probe device by emitting a red light if the device is too tilted to far up or a green light when the orientation of the device is acceptable Secondly the indicator light color chan...

Страница 8: ...n STEP 2 You may choose to annotate the measurement result with the eye s side right left information Note that the device s default selection is NO eye side information Select between the OD right ey...

Страница 9: ...17 8 17 4 18 2 mHg ME2 RECEI IF PRIN CONNE PRIN PRINTIN NOT FOUND PRINTER 12 PRINTER MENU If Printer is not found mmHg 12 6 OS 18 09 2017 07 37 45 AM 18 09 2017 07 37 45 AM 2 100 TO MENU 1 6 EYE OD O...

Страница 10: ...successfully made from one eye you may measure the IOP in the other eye or make a repeat measurement in the same eye by repeating the steps 1 5 above The device does not automatically switch from one...

Страница 11: ...2 6 3 6 4 6 5 6 0 LI LI LI LI 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU ME21 PAIR ME21 TEST ME21 UNPAIR BACK ME21 PAIRED TEST PRINTING REMOVED CONNECTION If PC Mobile Device is not found nne...

Страница 12: ...11 1 12 1 13 1 15 12 0 12 2 12 3 12 4 12 6 12 4 12 6 12 8 12 8 12 5 12 1 IF PRINTER CONNECTED 1 14 1 16 2 2 2 3 1 17 1 18 1 8 1 7 3 2 3 3 17 6 23 9 20 8 16 5 17 6 17 4 23 8 1 4 1 5 EYE EYE OFF PATIEN...

Страница 13: ...D 1 PATIENT ID ID02 PATIENT ID 5 1 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU If Printer is connected ME21 4 5 TO MENU BLUETOOTH Icare EXPORT SOUND SOUND SOUND SOUND SOUND SOUND OFF 12 2 12 3...

Страница 14: ...4 14 6 14 3 14 5 14 7 14 9 14 8 14 10 14 11 14 ERRORs ALERTs 14 13 13 8 13 9 SERVICE 14 12 INFO 9 0 DATE 16 03 2017 TIME 17 15 10 0 11 0 10 1 10 2 10 3 8 0 8 1 8 3 8 4 LANGUAGE ENGLISH 8 2 1 2 3 4 BR...

Страница 15: ...OAD LOAD mmHg mmHg mmHg mmHg mmHg mmHg 1 0 1 9 1 10 1 11 1 12 1 13 1 15 12 0 12 2 12 3 12 12 12 4 12 8 12 5 IF PRINTER CONNECTED 2 1 2 2 2 3 2 0 MEASURE 1 8 1 7 HISTORY mmHg 3 1 3 0 3 2 3 3 17 6 23 9...

Страница 16: ...on installed The Icare Export will transfer the data into the Icare CLINIC software Operators can then access the data by using the Icare CLINIC software with a web browser in an internet connected de...

Страница 17: ...s collar and probes are small objects and may be accidentally swallowed REPLACING THE PROBE BASE Change the probe base after six months If the error message Change is shown on the display screen more...

Страница 18: ...ailure Non ionizing radiation Manufacturer Do not dispose of this product with household waste Send to appropriate facility for recovery and recycling EU WEEE European Union Directive for Waste of Ele...

Страница 19: ...the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy...

Страница 20: ...utions regarding EMC and needs to be installed and put into service according to EMC information provided in following tables Portable and mobile RF communications equipment can affect Icare ic200 TA0...

Страница 21: ...terruption and voltage variations on power supply lines IEC 61000 4 11 0 UT for 0 5 cycle 1 phase 0 UT for 1 cycle 70 UT for 25 30 cycles 50 60 Hz 0 UT for 250 300 cycles 50 60 Hz NOT APPLICABLE NOT A...

Страница 22: ...vicinity of equipment marked with the following symbol GUIDANCE AND MANUFACTURER S DECLARATION IEC 60601 1 2 2014 EDITION 4 0 ELECTROMAGNETIC IMMUNITY Icare ic200 TA031 is intended for use in a profes...

Страница 23: ...ENGLISH Icare ic200 tonometer manual www icaretonometer com 23 NOTES All manuals and user guides at all guides com...

Отзывы: