ENGLISH
Icare ic200 tonometer manual
www.icaretonometer.com
20
ELECTROMAGNETIC DECLARATION
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating normally.
WARNING
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
PRECAUTION
The measurement method of the Icare ic200 tonometer is based on magnetic induction and therefore an external magnetic field in line with the probe may
prevent the measurement. In such case the tonometer will continuously ask to repeat the measurement. Situation can be solved either by removing the source
of interference from the vicinity of the device or by performing the measurement in different location with no such interference.
Icare ic200 (TA031) tonometer is a class B equipment and needs special precautions regarding EMC and needs to be installed and put into service according to
EMC information provided in following tables.
Portable and mobile RF communications equipment can affect Icare ic200 (TA031) Tonometer.
The Essential Performance of the Icare ic200 (TA031) tonometer is to measure the Intraocular Pressure (IOP) and to display
the measurement results.
GUIDANCE AND MANUFACTURER’S DECLARATION IEC 60601-1-2:2014; EDITION 4.0 - ELECTROMAGNETIC EMISSIONS
Icare ic200 (TA031) is intended for use in a professional healthcare environment with electromagnetic characteristics specified below.
The user of the Icare ic200 Tonometer (TA031) should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare ic200 (TA031) is battery operated and uses RF energy only for its internal function. Therefore,
its RF emissions are low and are not likely to cause any interference in nearby equipment
RF emissions CISPR 11
Class B
Icare ic200 (TA031) is suitable for use in all establishments, including domestic establishments and
those directly connected to public low-voltage power supply network that supplies buildings used
for domestic purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
NOT APPLICABLE
Voltage fluctuations flickering
emissions
IEC 61000-3-3
NOT APPLICABLE
NOT APPLICABLE