DESCRIPTION
Microprocessor controlled switching Power Supply
a
Last settings are automatically restored at power up.
a
IBI SCIENTIFIC Power Supplies are equipped with an automatic restart
system in case of power failure.
When the power returns, an audible alarm sounds for 10 seconds and the
power automatically restarts with the previous set values.
If during the power failure the electrophoresis unit is disconnected from the
power supply, the alarm still sounds when the power returns. Because no
load is connected the power supply will immediately shut-down and set
the output to zero.
a
Direct reading of programmed set values and actual values before and
during the cycle.
a
Volt and mA adjustable during a cycle.
a
Stabilization and automatic crossover between the parameters according
to the set limitation values and when output limits are reached.
2 LED indicate the constant mode.
a
2 operating modes: constant voltage - constant current.
a
2 LED Displays for set and output Volt, Current.
a
Settings by tactile switches on moisture resistant membrane panel.
a
Battery-backed memory feature "save" last output set values in the event of
a power failure or when the run is terminated and the unit is turned off.
a
4 recessed safety output jacks allow simultaneous operation of
4 electrophoresis units.
a
Fault detection and 500µA current leakage detection or no load automatically
shuts down output and indicates the fault by STOP LED blinking and audible
alarm.