SH-300 XL SPECIFICATIONS
Power supply voltage fluctuations not to exceed
± 10% of the normal voltage
Power supply, 110V model
90 - 130V; 50 - 60Hz; T4A fuses
Power supply, 220V model
180 - 260V; 50 - 60Hz; T2A fuses
Fuse value in the mains plug (UK)
3 A
Rated input power/current
150 VA
SH-300 XL
Volt range
1 - 300 Volts; 1Volt step
Current range
1 - 400 mA; 1mA step
Power
0,6 - 120Watts
Mode of operation
Continuous
Switching frequency
23 kHz
Output regulation stability
± 0,2% FS ± 1/2 digit
Minimum output
1 Volts; 15µA; 0,6 Watts
Value display accuracy
± 0,2 FS ± 1/2 digit
Power failure during the run
Audible alarm and automatic restart
with previous set values when power
returns
Fault detection
Output supply stop, audible alarm
STOP LED blinking
Fault status
Output to earth leakage
Output open circuit
Output short circuit
No regulation (Overheating,
power circuit fault)
Ground leakage
Detection level 500µA
Output to ground impedance
10M
Ω
min bypassed by 1nF max
Size
21 cm x 18 cm x 11 cm (D x W x H)
Weight
2 kg
Environmental conditions
Indoor use
Altitude up to 2000m
Temperature 10°C - 40°C
Maximum relative RH 80% for temperature up to 31°C decreasing linearly
to 50% RH at 40°C.
9