FAULT
A fault prevents the START mode operating or immediately cuts the power supply
during the gel run.
Simultaneously, the message
“FLT”
(fault) appears on the display, the audible
alarm rings and the led STOP blinks.
This automatic cut off indicates one of the following situations:
• Output to ground leakage,
• Output open circuit,
• Tank’s leads disconnected or defective,
• Output short circuit,
• Overload.
a
Press STOP to resume and to cease the audible alarm.
a
Look for the wrong situation.
a
Press START to run again.
a
If the unit returns to STOP mode, contact IBI SCIENTIFIC or your
supplier technical service.
AUTOMATIC RESTART
The power will automatically RESTART with the set values when the
power is operating again after a power failure or repetitive micro
failures during a cycle.
Simultaneously, the message
“FLT”
is displayed and the STOP LED
blinks for 10 seconds before operating again.
19