ibaDeskline SAS
Manual
Issue 1.4
9
4
Safety instructions
4.1
Designated use
The device may be used only for the following applications:
Machine test and commissioning of industrial systems
Measurement data logging and analysis
Applications of ibaSoftware products (ibaPDA, ibaLogic etc.)
4.2
Proper installation site
According to manufacturer's specifications the device can only be installed on an even
non-slip installation surface, for example table or rack.
Electric Shock!
Before opening the device disconnect it from the mains and pull out the mains plug from
the socket equipped with earthing contact!
Never use the device with a damaged mains cable!
Before opening the device disconnect it from the mains by unplugging the cable from
the socket equipped with an earthing contact and wait for several minutes until the com-
ponents have cooled down! Thus you will avoid injuries due to the electric shock or
burns!
Connect the device to one supply voltage only according to specifications on the built-
in power supply unit!
Always use a socket equipped with earthing contact! Use a terminal strip with overvolt-
age protection or an uninterruptible power supply (UPS)!
Always connect the device to earthed power networks (TN-networks according to VDE
0100 Part 300 and IEC 60364-3)! The operation via ungrounded networks or networks
(IT networks) earthed via impedance is not permitted!
In case of faults, a defective device or a possible defect pull out the mains plug from the
socket equipped with earthing contact!
Never put a damaged device into operation!
Pay attention to sharp edges in the housing!
Never switch off the device by means of the mains switch without previous shutdown of
the device.
Electrostatic discharge!
Touch the components only in electrostatically discharged state!