IACER Srl
79
MNPG178
COMPATIBILITY ELECTROMAGNETIC TABLES
Guidance and manufacturer’s declaration – ELECTROMAGNETIC EMISSIONS – FOR
ALL EQUIPMENT AND SYSTEMS
The I-TECH PHYSIO 4/EMG is intended for use in the electromagnetic environment
specified below. The customer or the user of the I-TECH PHYSIO 4/EMG should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment -
guide
RF emissions
CISPR 11
Group 1
The I-TECH PHYSIO 4/EMG uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The I-TECH PHYSIO 4/EMG is
suitable for domestic
establishment and in
establishment directly connected
to the public low-voltage power
supply network that supplis
buildings used for domestic
purposes.
Harmonics emissions
IEC 61000-3-2
Class A
Voltage
fluctuation/flicker
emissions
IEC 61000-3-3
Compliant
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY – FOR ALL
EQUIPMENT AND SYSTEMS
The I-TECH PHYSIO 4/EMG is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that it is used in
such an environment
Immunity test
Test level
IEC 60601
Compliance level
Electromagnetic
environment - guide
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV; +8kV in
contact
±8kV; +15kV on
air
±6kV; +8kV in
contact
±8kV; +15kV on air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
±2kV
for power
±2kV
for power supplies
Mains power quality
should be that of a typical