![I-Tech PHYSIO 4 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/i-tech/physio-4/physio-4_user-manual_3153423007.webp)
IACER Srl
3
MNPG178
The CE0068 marking on the device applies exclusively to medical programs
(see the following paragraphs concerning the detailed description of the
programs).
Technical features
Characteristics
Specifications
Power supply
Rechargeable batteries NiMH 7.2V, 2000mAh
(dimensions 6xAA)
Battery charger
Model EA1018G-12V
100-240VAC, 50-60Hz, 1.0A input
12VDC, 2.0A output
Not replaceable 250VAC, 2A fuse
Max. absorbed current 1.6 A
Absorption
DC 12V, 1.5A
Output power
120mA (with 1000 Ohm rated load)
Isolation class (EN
60601-1)
II
Applied part (EN 60601-
1)
BF
Dimensions (mm)
(length x width x height)
180x110x50
Protection level IP
IP20
Conditions of use
Environmental
temperature
From +10 to +40°C
Relative humidity
From 15 to 93%
Atmospheric pressure From 500 to 1060hPa
Storage and
transportation
conditions
Environmental
temperature
From +5 to +40°C
Relative humidity
From 15 to 93%
Atmospheric pressure From 500 to 1060hPa
Electrotherapy (models 4 and EMG)
Number of programs
A total of 86 programs divided as follows:
-
n. 26 prog.
TENS/ANALGESIC/IONOPHORESIS
(diadynamic, faradic, Kotz and interferential
currents).
-
n. 25 prog. REHAB (denervated muscles,
Содержание PHYSIO 4