IACER Srl
30
MNPG178
W19 •
CP Current – Short period (medical program)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
It is composed of alternated single-phase (3 seconds) and biphasic (3
seconds) waveforms. The short term current has an essentially
dynamogenic effect. This action promotes contraction of the striated
muscles, improves the nutritional status of the tissue and facilitates the re-
absorption of post-traumatic oedemas. Frequency alternating current is
clearly perceptible: the DF current produces a light tremble while the MF
current produces a strong pulsation.
Physiological effects: This kind of current has a strong dynamogenic effect
(or trophic effect) and a great analgesic effect, especially in the case of
chronic pain. Moreover, the CP current is particularly effective with regard
to the re-absorption of haematomas and oedemas.
Indications: It is used in the short term to treat pain resulting from the
inflammation of tendons, articular capsule and soft tissues (tendinitis,
bursitis, rheumatoid arthritis and trauma in general). Session duration: 15
minutes.
Electrodes’ positioning: form a square above the area to be treated as
shown in
Figure 2
.
Intensity: to be adjusted between the threshold of perception and pain.
W20 • LP Current – Long Period (medical program)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
It is composed of alternated single-phase (10 seconds) and biphasic (5
seconds) waveforms. The long term current has essentially an inhibiting
action on muscle and sensibility: as a consequence it produces analgesia
and relaxation of the striated muscles.
Session duration: 15 minutes.
Electrodes’ positioning: : form a square above the area to be treated as
shown in
Figure 2
.
Intensity: to be adjusted between the threshold of perception and pain.
W21 • HVPC – High Voltage Pulsed Current (medical program)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
HVPC can be used for the reconstruction of tissue, the re-absorption of
oedemas and the reduction of pain.
For repairing tissue, place a sterile gauze dampened with a saline solution
on the wound: if the anode (positive electrode) is placed over the wound,
Содержание PHYSIO 4