MNPG178
Parameter description
Minimum
Maximum
Program total time
1 minute
60 minutes
M10/M11 • EMS ramp of 2 seconds (non-medical programs)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
Table of settable values:
Parameter description
Minimum
Maximum
Contraction frequency
20 Hertz
120 Hertz
Contraction period
4 seconds
10 seconds
Recovery frequency
0 Hertz
12 Hertz
Recovery period
2 seconds
26 seconds
Impulse width
60 microseconds
500 microseconds
Program total time
1 minutes
60 minutes
M12/M13 • EMS ramp of 3 seconds (non-medical programs)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
Table of settable values:
Parameter description
Minimum
Maximum
Contraction frequency
20 Hertz
120 Hertz
Contraction period
6 seconds
10 seconds
Recovery frequency
0 Hertz
12 Hertz
Recovery period
2 seconds
24 seconds
Impulse width
60 microseconds 500 microseconds
Program total time
1 minute
60 minutes
M14/M15 • FES (non-medical program)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
Available programs to set treatments for incontinence prevention and
cure.
Table of settable values:
Parameter description
Minimum
Maximum
Contraction frequency
20 Hertz
120 Hertz
Contraction period
6 seconds
10 seconds
Recovery frequency
0 Hertz
12 Hertz
Recovery period
2 seconds
24 seconds
Impulse width
60 microseconds 500 microseconds
Program total time
1 minute
60 minutes