I.A.C.E.R. Srl
17
MNPG273-00
Vertebral column
Lumbar
Suggestions for correct use:
•
The device and the applicators are designed to operate in the
temperature ranges indicated with treatments of up to 12
consecutive hours;
•
In the P1-P41 programs, if you set an intensity higher than 100,
with extended therapy times, the 3-solenoid belt could heat up
thereby making the therapy uncomfortable: it is advisable to
split up the treatment sessions and do not exceed 2/3
consecutive hours of therapy;
•
In P42-P45 programs, if you want to set an intensity higher than
100 with treatments over 2 hours, we recommend using the
professional applicator with a pair of solenoids available as an
optional accessory for MAG3000.
Instructions for using the preset programs
To use
MAG3000
by freely setting the parameters regarding therapy
time and magnetic field intensity, follow these simple steps:
1. Connect the applicator (or applicators) to the device by connecting
the plug of the applicator cable to one of the two ports (CH1-CH2)
on the panel at the top of the appliance;
2. Connect the mains cable to the power supply and then connect the
power supply outlet to the circular connector on the panel at the top
of the device, near the ON/OFF switch;
3. Connect the plug of the mains cable to the mains socket (110-
230VAC, 50-60 Hz);
4. Press the ON/OFF switch on the small panel in the upper part so it
is in the ON position: the display will show the I-TECH MEDICAL
DIVISION logo and then the main menu screen;
5. Select "Single Patient" and scroll through the programs using the
and
buttons and select the desired program