I.A.C.E.R. Srl
34
MNPG273-00
RF immunity aspects
The MAG3000 is intended to work in the electromagnetic environment specified below.
The customer or user should make sure that it is used in such environment
Immunity test Test level EN
60601-1-2
Compliance
level
Electromagnetic environment -
guidance
Conduct RF
EN 61000-4-6
3 Veff. 150kHz
to 80MHz
3 Veff. 150kHz
to 80MHz
Portable and mobile RF
communications equipment should not
be used near any part of the
equipment, including cables, except
when respecting the recommended
separation distances calculated from
the equation applicable to the
transmitter frequency
Recommended separation distances
d = 1.2
⋅√
P 150kHz to 80MHz
d = 1.2
⋅√
P 80 MHz to 800 MHz
d = 2.3
⋅√
P 800 MHz to 2.7 GHz
where P is the maximum output
power of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters
(m).
Radiated RF
EN 61000-4-3
10 V/m 80 MHz
to 2.7 GHz
10 V/m 80 MHz
to 2.7 GHz
The field strength from fixed RF transmitters, as determined by an
electromagnetic site survey, may be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Recommended separation distance between portable and mobile radio
communication devices and the MAG3000 device
The MAG3000 is intended to operate in an electromagnetic environment in which RF irradiated
disturbances are controlled. The customer or the operator of the device can help prevent
electromagnetic interference by ensuring a minimum distance between mobile and portable RF
communications devices (transmitters) and the device, as recommended below, in relation to
the maximum output power of the radio communication devices.
Maximum rated
output power of
the transmitter (W)
Separation distance to the frequency of the transmitter (m)
150kHz to 80MHz
d = 1.2
⋅√
P
80MHz to 800MHz
d = 1.2
⋅√
P
800MHz to
2.7GHz
d = 2.3
⋅√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Содержание MAG3000