Immunity aspects
The MAG3000 is intended to work in the electromagnetic environment specified below.
The customer or user should make sure that it is used in such environment
Immunity test
Test level EN
60601-1-2
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
±
8kV contact
±
15kV air
±
8kV contact
±
15kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
EN 61000-4-4
±
2kV power
supply lines
±
2kV power
supply lines
Mains power quality should
be that of a typical business
or hospital environment.
Impulses
EN 61000-4-5
±
1kV differential
mode
±
1kV differential
mode
Mains power quality should
be that of a typical business
or hospital environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
EN 61000-4-11
0%
UT
for 0.5 cycles,
different angles
0% U
T
for 1 cycle
70% U
T
for 25/30 cycles
0% U
T
for 5 seconds
0% U
T
for 0.5 cycles,
different angles
0% U
T
for 1 cycle
70%
UT
for 25/30 cycles
0% U
T
for 5 seconds
Mains power quality should
be that of a typical business
or hospital environment. If
the user requires continuous
operation even during the
interruption of the mains
voltage, it is recommended
to power the device with an
uninterruptible power supply
(UPS) or with batteries.
Magnetic field at
mains frequency
EN 61000-4-8
30 A/m
30 A/m
The magnetic fields at mains
frequency should be at levels
typical of a business or
hospital environment.