Guidance and manufacturer's declaration - ELECTROMAGNETIC
IMMUNITY - FOR ALL EQUIPMENT AND SYSTEMS
I-PRESS is designed to work in the electromagnetic environment specified
below. The customer or the user of I-PRESS must ensure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
(>95% dip in
U
T
) for 5s
(>95% dip in
U
T
) for 5s
Magnetic field at
power frequency
(50/60Hz)
IEC 61000-4-8
30A/m
30A/m
The magnetic fields at
power frequency
should be at levels
typical of a commercial
or hospital
environment.
Note: UT is the AC mains voltage prior to application of the test level.