Guidance and manufacturer's declaration - ELECTROMAGNETIC
IMMUNITY - FOR EQUIPMENT AND SYSTEMS NOT SUPPORTING VITAL
FUNCTIONS
I-PRESS is designed to work in the electromagnetic environment specified
below. The customer or the user of I-PRESS must ensure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Portable and mobile RF communications equipment should not be used
near any part of the device, including cables, except when respecting the
recommended separation distances calculated from the equation
applicable to the transmitter frequency.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3V
eff
from 150kHz
to 80MHz
6V
eff
from 150kHz
to 80MHz for
ISM band
3V
eff
([
V1
] V)
6V
eff
([
V1
] V)
=
=
P
V
d
1
5
,
3
P
V
d
=
1
12
for ISM band
Radiated RF
IEC 61000-4-3
3V/m
from 80MHz
to 2.7GHz
3V/m
[
E
1
] V/m
P
E
d
=
1
12
from 80MHz to 800MHz
P
E
d
=
1
7
from 800MHz to 2.7GHz
Radiated RF for
radio
communication
devices
IEC 61000-4-3
3V/m
from 80MHz
to 6GHz
3V/m
[
E
1
] V/m
P
E
d
=
1
6
from 80MHz to 6GHz