IACER Srl
13
MNPG338-00
1.1.1.
Attach the connector relating to the tube with 8 endings (for
using two leg cuffs) or the connector relating to the tube with 4
endings if using a single leg cuff;
1.1.2.
Put on the leg cuffs (or the single leg cuff) and close the zip up
to the top, then close the Velcro fastener; if it is necessary to
increase the circumference of the leg cuffs, add the extenders
for leg cuffs available on request;
1.1.3.
Insert the pressure orthotics at the lower ends of the leg cuffs,
under the sole of the foot.
N.B.:
it is recommended to cut the non-woven fabric band
according to the shape of the pressure orthotics and place it
between the foot and the pressure orthotics.
1.1.4.
When wearing the leg cuffs make sure that the connection
inlets of the tubes is always on the inside of each leg cuff; if only
one leg cuff is used, the position of the connection inlets does
not matter.
1.1.5.
Connect the tubes to the 4 inlets on each leg cuff, making sure
to connect the longer tube to the topmost connector (darker in
colour) on the leg cuff (thigh).
1.2.
Arm cuff
1.2.1.
Put on the arm cuff and connect the tubes to the inlets on the
arm cuff itself, making sure to connect the longer tube to the
topmost connector (darker in colour) on the arm cuff;
CONNECTION
INLETS
VELCRO
ZIP