74
i-SmartCare 10 Blood Gas Analyzer
Blood Sample
,
continued
Sample Results,
continued
4. Sample Analysis
5. If there are configured
Critical Limit
ranges, and when the sample
result is higher than the
Critical Limit
,
appears.
appears if the sample result is lower than the
Critical Limit
.
6. When the sample result is out of the measuring range, no result is
shown. And "
Reportable Range Error
" message appears.
7. The result with the Slope Error does not produce the result values and
“
Slope Error
” message appears.
8. If
Calculated
tab is chosen, the calculated output using formulae is
produced.
Note:
Any value that cannot be calculated under
Calculated
tab appears
as blank along with the
Incalculable
message.
Содержание i-SmartCare 10
Страница 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Страница 3: ......
Страница 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Страница 5: ......
Страница 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Страница 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Страница 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Страница 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Страница 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Страница 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Страница 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...